首页> 外文期刊>Anesthesia and Analgesia: Journal of the International Anesthesia Research Society >Intrathecal morphine pump malfunction due to leakage at the catheter connection site: a rare problem and its prevention.
【24h】

Intrathecal morphine pump malfunction due to leakage at the catheter connection site: a rare problem and its prevention.

机译:鞘内吗啡泵由于导管连接部位的泄漏而发生故障:这是一个罕见的问题及其预防措施。

获取原文
获取原文并翻译 | 示例
           

摘要

Implanted intrathecal drug delivery systems provide excellent pain relief and reduced drug toxic-ity in patients with refractory cancer pain.1'2 Common complications include postpuncture headache (15%), external leakage of cerebrospinal fluid (3.5%), CSF hygroma (1.5%), catheter tip dislodgement (1.5%) and catheter system leakage (1.5%). Infection of implanted systems are a rare occurrence, mostly in the first 3 mo after implantation.3 We report a pump malfunction resulting from pump-catheter connection failure and a strategy to avoid this complication. A 72-yr-old patient with bladder cancer meta-static to the sacrum and lumbar plexus was receiving the equivalent of 750 mg morphine per day postop-eratively, but with significant sedation and still inadequate pain control. A percutaneous intrathecal catheter was placed and an infusion of morphine begun. The patient improved rapidly, showing much lower pain scores while being alert and mobile.
机译:植入的鞘内给药系统可为顽固性癌症疼痛患者提供出色的疼痛缓解和降低的药物毒性。1'2常见并发症包括穿刺后头痛(15%),脑脊液外漏(3.5%),CSF潮气(1.5%) ),导管尖端移位(1.5%)和导管系统泄漏(1.5%)。植入系统的感染很少发生,主要是在植入后的最初3个月。3我们报告了由泵导管连接故障引起的泵故障,以及避免这种并发症的策略。一名患有to骨和腰丛转移性转移的膀胱癌的72岁患者,术后每天接受相当于750 mg吗啡的治疗,但镇静作用明显,并且疼痛控制仍不充分。放置经皮鞘内导管,开始输注吗啡。该患者迅速好转,在保持警觉和移动的同时,疼痛评分大大降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号