Infrared Thermographic Assessment of Cooling Effectiveness in Selected Dental Implant Systems

机译:选定牙科系统中的冷却效能的红外线热量分析

获取原文
获取原文并翻译 | 示例

摘要

The excessive temperature fluctuations during dental implant site preparation may affect the process of bone-implant osseointe-gration. In the presented studies, we aimed to assess the quality of cooling during the use of 3 different dental implant systems (BEGO?, NEO BIOTECH?, and BIOMET 3i?), The swine rib was chosen as a study model. The preparation of dental implant site was performed with the use of 3 different speeds of rotation (800,1,200, and 1,500 rpm) and three types of cooling: with saline solution at room temperature, with saline solution cooled down to 3°C, and without cooling. A statistically significant difference in temperature fluctuations was observed between BEGO and NEO BIOTECH dental systems when cooling with saline solution at 3°C was used (22.3°C versus 21.8°C). In case of all three evaluated dental implant systems, the highest temperature fluctuations occurred when pilot drills were used for implant site preparation. The critical temperature, defined in the available literature, was exceeded only in case of pilot drills (of all 3 systems) used at rotation speed of 1,500 rpm without cooling.
机译:牙科植入部位制剂期间的过度温度波动可能会影响骨植入骨膜常规的过程。在本研究中,我们旨在评估使用3种不同的牙科植入系统(BEGO?,Neo Biotech?,和Biomet 3i的冷却质量,选择猪肋作为研究模型。使用3种不同的旋转速度(800,1,200和1,500rpm)和三种冷却剂进行牙科植入部位的制备:在室温下用盐水溶液,盐水溶液冷却至3°C,没有冷却。在使用3℃的盐水溶液(22.3°C与21.8°C)之间使用盐水溶液冷却时,在BEGO和Neo Biotech牙科系统之间观察到统计学上的显着差异。在所有三种评估的牙科植入系统的情况下,当飞行钻头用于植入部位准备时发生最高温度波动。仅在可用文献中定义的临界温度仅超过500 rpm的试验钻头(所有3个系统)而无需冷却。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号