Validation of error analyzer in surgical navigation systems using 3-D position sensor

机译:使用3-D位置传感器验证手术导航系统中的错误分析仪

获取原文
获取原文并翻译 | 示例
           

摘要

Error analysis of surgical navigation systems, which integrate the real space with the virtual one using 3-D position sensor, has been studied in order to estimate the system accuracy. Experiments using phantoms need much time and sometimes cannot specify the error factor. Thus, the accuracy validation tool with the computer simulation, which is called "registration graph", has been proposed. We have extended it so that the graph includes the information of the system design in order to estimate the system accuracy more easily. To validate reliability of our error analysis tool, we compared the accuracy estimated by the tool to actual one in three systems.
机译:研究了手术导航系统的误差分析,该系统与使用3-D位置传感器的虚拟体集成了真实空间,以估计系统精度。 使用幻像的实验需要多长时间,有时不能指定错误因子。 因此,已经提出了具有计算机仿真的精度验证工具,该工具被称为“注册图”。 我们已经扩展了它,以便该图包括系统设计的信息,以便更容易估计系统精度。 为了验证我们的错误分析工具的可靠性,我们将工具估计的准确性与三个系统中的实际估计进行了比较。