首页> 外文期刊>Anaesthesia: Journal of the Association of Anaesthetists of Great Britain and Ireland >Evaluation of a new analyser for rapid measurement of blood propofol concentration during cardiac surgery
【24h】

Evaluation of a new analyser for rapid measurement of blood propofol concentration during cardiac surgery

机译:评估用于心脏手术期间快速测量血中丙泊酚浓度的新型分析仪

获取原文
获取原文并翻译 | 示例

摘要

We report laboratory and clinical evaluations of a blood propofol concentration analyser. Laboratory experiments used volunteer blood spiked with known propofol concentrations over the clinically relevant concentrations from 0.5 to 16 μg.ml -1 to assess linearity and the influence of haematocrit and concurrent drug administration. Analyser concentrations demonstrated excellent linearity (R 2 = 0.999). Blood spiked with commonly used drugs showed no significant variation compared to unspiked controls. Propofol measurements were largely independent of haemoglobin concentration. A 6% decay in propofol concentration was observed at the highest prepared concentration. Clinical performance of the analyser was assessed using 80 arterial blood samples from 72 patients receiving propofol infusions during cardiac surgery. Samples were processed using the propofol analyser, and high performance liquid chromatography (HPLC) used as a gold-standard comparator. These data demonstrated excellent agreement between the propofol analyser and HPLC with a bias of 0.13 μg.ml -1 and precision of -0.16 to 0.42 μg.ml -1.
机译:我们报告了血液和异丙酚浓度分析仪的实验室和临床评估。实验室实验使用自愿性血液加标有已知丙泊酚浓度(临床上相关浓度为0.5至16μg.ml-1)来评估线性度,血细胞比容和同时给药的影响。分析仪浓度显示出极好的线性(R 2 = 0.999)。与未加标的对照组相比,加有常用药物的加标血液显示无明显变化。异丙酚的测定很大程度上与血红蛋白浓度无关。在最高制备浓度下,丙泊酚浓度下降了6%。使用来自心脏手术期间接受丙泊酚输注的72位患者的80份动脉血样本评估了分析仪的临床性能。使用丙泊酚分析仪处理样品,并使用高效液相色谱(HPLC)作为金标准比较剂。这些数据表明丙泊酚分析仪与HPLC之间具有极好的一致性,偏差为0.13μg.ml-1,精度为-0.16至0.42μg.ml-1。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号