首页> 外文期刊>Anaesthesia: Journal of the Association of Anaesthetists of Great Britain and Ireland >Effects of laryngoscope handle light source on the light intensity from disposable laryngoscope blades.
【24h】

Effects of laryngoscope handle light source on the light intensity from disposable laryngoscope blades.

机译:喉镜手柄光源对一次性喉镜叶片光强度的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A bench-top study was performed to assess the effects of different laryngoscope handles on the light intensity delivered from disposable metal or plastic laryngoscope blades. The light intensity from both the handle light sources themselves and the combined handle and laryngoscope blade sets was measured using a custom-designed testing system and light meter. Five samples of each disposable blade type were tested and compared with a standard re-usable stainless steel blade using three different handle/light sources (Vital Signs LED, Heine 2.5?V Xenon and 3.5?V Xenon). The light intensity delivered by the disposable blades ranged from 790 to 3846 lux for the different handle types. Overall, the 3.5?V Heine handle delivered the highest light output (p?
机译:进行了一项台式研究,以评估不同喉镜手柄对从一次性金属或塑料喉镜刀片发出的光强度的影响。使用定制设计的测试系统和照度计测量来自手柄光源本身以及组合的手柄和喉镜刀片组的光强度。测试了每种一次性刀片的五个样品,并与使用三种不同手柄/光源(生命体征LED,Heine 2.5?V氙气和3.5?V氙气)的标准可重复使用的不锈钢刀片进行了比较。对于不同的手柄类型,一次性刀片提供的光强度范围为790至3846 lux。总体而言,与其他手柄相比,3.5?V的Heine手柄可提供最高的光输出(p 0.007)。对于一次性刀片,塑料的总光输出明显高于金属刀片(p≤0.001)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号