The laboratory performance of the enFLOW (R), buddy lite (TM) and ThermoSens (R) fluid warmers

机译:enFLOW(R),buddy lite(TM)和ThermoSens(R)液体加热器的实验室性能

获取原文
获取原文并翻译 | 示例
           

摘要

We measured heating of isotonic saline by three fluid warmers in six experiments: saline at 5 degrees C or 20 degrees C delivered at 30, 50 or 100ml.min(-1). At the three flow rates, the enFLOW((R)), buddy lite and ThermoSens((R)) systems heated 5 degrees C saline to mean (SD) temperatures of: 41.1 (0.5) degrees C, 37.7 (0.6) degrees C and 39.1 (0.6)degrees C; to 40.3 (0.8) degrees C, 33.9 (1.6) degrees C and 39.3 (0.7)degrees C; and to 37.1 (0.8) degrees C, 24.0 (1.3) degrees C and 37.6 (1.0)degrees C, respectively, p<0.0001 for each experiment. The mean (SD) times taken to heat 5 degrees C saline were: 16.6 (1.7) s, 258.4 (58.9) s and 134.2 (79.6)s; 16.9 (1.8) s, 256.2 (62.2) s and 182.5 (74.5)s; and 21.5 (1.5) s, 275.9 (49.3) s and 313.5 (18.0)s, respectively, p<0.0003 for each experiment. The results for saline at 20 degrees C were similar. The enFLOW system heated saline above 36 degrees C faster than the ThermoSens system, whereas the buddy lite often failed to achieve 36 degrees C.
机译:我们在六个实验中通过三个液体加热器测量了等渗盐水的加热情况:以30、50或100ml.min(-1)输送的5摄氏度或20摄氏度盐水。在这三个流速下,enFLOW(R),buddy lite和ThermoSens(R)系统将5摄氏度的盐水加热至41.1(0.5)摄氏度,37.7(0.6)摄氏度的平均(SD)温度。和39.1(0.6)摄氏度;至40.3(0.8)摄氏度,33.9(1.6)摄氏度和39.3(0.7)摄氏度;对于每个实验,分别达到37.1(0.8)摄氏度,24.0(1.3)摄氏度和37.6(1.0)摄氏度,p <0.0001。加热5摄氏度盐水所需的平均(SD)时间为:16.6(1.7)s,258.4(58.9)s和134.2(79.6)s; 16.9(1.8)秒,256.2(62.2)秒和182.5(74.5)秒;和21.5(1.5)s,275.9(49.3)s和313.5(18.0)s,每个实验的p <0.0003。在20摄氏度下盐水的结果相似。 enFLOW系统将盐溶液加热到36摄氏度以上的速度比ThermoSens系统快,而伙伴lite常常无法达到36摄氏度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号