Ventilation failure as a result of internal battery leakage in an Ulco Ev500 ventilator.

机译:Ulco Ev500呼吸机内部电池泄漏导致的通风故障。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The provision of adequate ventilation is crucial to the safe conduct of anaesthesia. We present a case of partial ventilator failure due to the destruction of ventilator components as a result of leakage of acid from the ventilator's internal battery. A 53-year-old man with a recent history of myocardial infarction and cardiogenic shock was anaesthetised for the insertion of an automatic implanted cardioverter defibrillator. Following induction and paralysis, he was intubated and intermittent positive pressure was maintained on an Ulco EV500 ventilator (Ulco Medical, Marrickville NSW) with a tidal volume of 500 ml and a peak inspiratory pressure of 18 cmH_2O.
机译:提供足够的通风对于安全进行麻醉至关重要。我们介绍了由于呼吸机内部电池漏酸导致呼吸机部件损坏而导致的部分呼吸机故障。一名患有心肌梗塞和心源性休克病史的53岁男子因插入自动植入的心脏复律除颤器而被麻醉。诱导和麻痹后,将其插管,并在Ulco EV500呼吸机(Ulco Medical,Marrickville NSW)上维持间歇性正压,潮气量为500 ml,吸气峰值为18 cmH_2O。