首页> 外文期刊>Anaesthesia and intensive care >Target controlled infusion pump failure due to worn drive nut.
【24h】

Target controlled infusion pump failure due to worn drive nut.

机译:由于驱动螺母磨损,导致目标控制的输液泵故障。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

During anaesthesia of a 67-year-old male undergoing cervical laminectomy at C3/4, an Alaris PK Mk III pump (CareFusion, Basingstoke, UK) delivering remifentanil alarmed 'PL3' when the infusion was recommenced after a new syringe was fitted. The pump would not recommence target- controlled infusion (TCI), necessitating a change to a manual infusion.
机译:在一名67岁男性在C3 / 4接受宫颈椎板切除术的麻醉期间,当安装了新注射器后建议重新输注时,提供瑞芬太尼的Alaris PK Mk III泵(CareFusion,英国贝辛斯托克)。泵不会重新启动目标控制的输液(TCI),因此需要更改为手动输液。