Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure

机译:使用NT-530P非接触式Tono /测厚仪进行测量的可重复性和可重复性以及中央角膜厚度与眼压的相关性

获取原文
获取原文并翻译 | 示例

摘要

Purpose. To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP. Methods. Forty-six eyes of healthy volunteers were measured by two examiners. Three consecutive measurements per eye were performed. Repeatability was assessed using the coefficient of variation, and reproducibility was assessed using Bland-Altman plots. Linear correlations were used to determine agreement between CCT and noncorrected IOP and CCT and corrected IOP, which was calculated using a formula built into the NT-530P. Results. The coefficient of variation for IOP was 6.4% and for CCT was 0.4%. The 95% limits of agreement between examiners were -0.17 ± 1.42 mmHg (range: -2.95 to 2.61 mmHg) for IOP, -0.93 ± 4.37 mum (range: -9.50 to 7.64 mum) for CCT. The corrected IOP was significantly higher than the noncorrected IOP (P = 0.010.3). The noncorrected IOP significantly correlated with CCT (r = -0.4883, P = 0.0006). The corrected IOP showed no significant correlation with CCT (r = -0.0285, P = 0.8509). Conclusions. NT-530P offered repeatability and reproducibility in both IOP and CCT measurements. The corrected IOP calculated using the NT-530P was independent of the CCT, suggesting that this IOP may be less influenced by the central corneal thickness.
机译:目的。调查使用非接触式Tono /测厚仪(NT-530P)进行眼内压(IOP)和角膜中央厚度(CCT)测量的可重复性和重现性,并评估CCT与IOP的相关性。方法。由两名检查员测量了健康志愿者的四十六只眼睛。每只眼睛进行三个连续测量。使用变异系数评估可重复性,并使用Bland-Altman图评估可重复性。线性相关用于确定CCT与未校正的IOP以及CCT与校正的IOP之间的一致性,这是使用NT-530P内置的公式计算的。结果。 IOP的变异系数为6.4%,CCT的变异系数为0.4%。考官之间协议的95%限制对于IOP为-0.17±1.42 mmHg(范围:-2.95至2.61 mmHg),对于CCT为-0.93±4.37 mum(范围:-9.50至7.64 mum)。校正后的IOP明显高于未校正的IOP(P = 0.010.3)。未经校正的IOP与CCT显着相关(r = -0.4883,P = 0.0006)。校正后的IOP与CCT无显着相关性(r = -0.0285,P = 0.8509)。结论NT-530P在IOP和CCT测量中均提供了可重复性和可重复性。使用NT-530P计算的校正后的IOP与CCT无关,这表明该IOP可能不受中央角膜厚度的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号