首页> 外文期刊>American Journal of Ophthalmology: The International Journal of Ophthalmology >Tear meniscus evaluation by anterior segment swept-source optical coherence tomography
【24h】

Tear meniscus evaluation by anterior segment swept-source optical coherence tomography

机译:前段扫频光学相干断层扫描评估泪液半月板

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To investigate the repeatability of tear meniscus measurements using anterior segment swept-source optical coherence tomography (OCT) and the relationship of tear meniscus measurements with tear film breakup time and Schirmer test results. Design: Prospective, observational, cross-sectional study. Methods: We enrolled 26 healthy subjects (26 eyes; 20 men and 6 women; mean age, 36.5 ± 6.8 years) at the University of Tokyo School of Medicine, Tokyo, Japan, in this study. Examinations were conducted in the following sequence: anterior segment vertical raster scans by swept-source OCT, tear film breakup time, and Schirmer test. Each subject underwent OCT measurements twice by the same grader, and the central upper and lower tear menisci height and area and the lower tear meniscus volume were examined. Each OCT image was evaluated by 2 masked graders using the software calipers. Results: The average upper and lower tear meniscus heights were 231 ± 78 μm and 256 ± 57 μm, respectively, and the average upper and lower tear meniscus areas were 18 829 ± 7823 μm2 and 21 903 ± 8173 μm2, respectively. The average tear meniscus volume was 0.1327 ± 0.051 mm3. The intergrader intraclass correlations for all the parameters were more than 95%. The OCT tear meniscus measurements and the Schirmer test scores were correlated significantly (P .05, Spearman nonparametric correlation analysis). However, tear film breakup time was not correlated significantly with any of the parameters of tear menisci (P .05). Conclusions: Anterior segment swept-source OCT is a noninvasive and practical method that can be used for quantitative evaluation of tear fluid.
机译:目的:探讨使用前节扫描源光学相干断层扫描(OCT)进行的泪液弯月面测量的可重复性,以及泪液弯月面测量与泪膜破裂时间和Schirmer测试结果之间的关系。设计:前瞻性,观察性,横断面研究。方法:本研究招募了日本东京大学东京大学医学院的26名健康受试者(26眼; 20名男性和6名女性;平均年龄36.5±6.8岁)。按以下顺序进行检查:通过扫频OCT进行前节垂直栅格扫描,泪膜破裂时间和Schirmer测试。每名受试者均由同一评分者两次进行OCT测量,并检查中央上,下泪液弯液面的高度和面积以及下泪液弯液面的体积。每个OCT图像均由2个蒙版的评分员使用软件卡尺进行评估。结果:平均上下撕裂弯液面高度分别为231±78μm和256±57μm,平均上下撕裂弯液面面积分别为18 829±7823μm2和21 903±8173μm2。平均撕裂弯月面容积为0.1327±0.051 mm3。所有参数的等级间等级内相关性均超过95%。 OCT泪液半月板测量值与Schirmer测试评分显着相关(P <.05,Spearman非参数相关分析)。然而,泪膜破裂时间与泪液半月板的任何参数均无显着相关性(P> .05)。结论:前节扫频OCT是一种无创且实用的方法,可用于定量评估泪液。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号