Effects of chemical restraint on electroretinograms recorded sequentially in awake, sedated, and anesthetized dogs

机译:化学限制对清醒,镇静和麻醉犬连续记录的视网膜电图的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Objective—To quantitatively and qualitatively compare electroretinography (ERG) recordings in awake, sedated, and anesthetized dogs.Animals—Six 6-month-old Beagles.Procedures—A brief ERG protocol for dogs was used. Following 1-minute and subsequent 5-minute dark adaptation, mixed rod-cone responses were recorded bilaterally with a handheld multispecies ERG device with dogs in each of 3 states of consciousness: awake, sedated (dexmedetomidine and butorphanol), and anesthetized (atropine and hydromor-phone, followed by propofol and midazolam and anesthetic maintenance with isoflurane). Low- and high-frequency noise levels were quantified via Fourier analysis, and the effect of consciousness state on signal amplitude, implicit time, and noise was analyzed via repeated-measures ANOVA. In addition, 13 veterinary ophthalmologists who were unaware of the dogs' consciousness states subjectively graded the ERG recordingquality, and scores for each tracing were compared.Results—ERG amplitudes were highest in awake dogs and lowest in anesthetized dogs. Implicit times were shortest in awake dogs and longest in anesthetized dogs. Differences in b-wave amplitudes and a-wave implicit times were significant. Neither low- norhigh-frequency noise levels differed significantly among consciousness states. Furthermore, no significant differences were identified among observers' scores assigned to ERG tracings.Conclusions and Clinical Relevance—Anesthesia and sedation resulted in significant attenuation and delay of ERG responses in dogs. Chemical restraint of dogs had no consistently significant effect on low- or high-frequency noise levels or on observer perception of signal quality.
机译:目的-定量,定性比较清醒,镇静和麻醉犬的视网膜电图(ERG)记录。动物-六只6个月大的比格犬。程序-使用简短的ERG犬规程。在进行1分钟和5分钟的黑暗适应后,用手持式多物种ERG设备双向记录混合的棒锥反应,该设备带有狗,处于以下三种意识状态:清醒,镇静(右美托咪定和布托啡诺)和麻醉(阿托品和静液电话,然后是异丙酚和咪达唑仑,并用异氟烷进行麻醉维持)。低频和高频噪声水平通过傅立叶分析进行量化,意识状态对信号幅度,隐式时间和噪声的影响通过重复测量方差分析进行分析。另外,对13名没有意识到狗的意识状态的兽医眼科医生主观地对ERG记录质量进行了评分,并比较了每种描记的得分。结果-清醒狗的ERG幅度最高,而麻醉狗的ERG幅度最低。隐性时间在清醒的狗中最短,在麻醉的狗中最长。 b波振幅和a波隐含时间的差异很明显。意识状态之间的低频或高频噪声水平均无显着差异。此外,在分配给ERG描记的观察者分数之间没有发现显着差异。结论和临床意义-麻醉和镇静会导致狗的ERG反应明显减弱和延迟。对狗的化学约束对低频或高频噪声水平或观察者对信号质量的感知均没有持续的显着影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号