Pressure gradient and inspiratory times required for valve opening of various holding chambers.

机译:各种保持腔室阀门打开所需的压力梯度和吸气时间。

获取原文
获取原文并翻译 | 示例
           

摘要

This study was designed to determine whether infants generate sufficient inspiratory pressure gradient to open the unidirectional valve of various commercially available holding chambers, as well as the time required to reach this gradient. A prospective, cross-sectional study was performed of 80 healthy infants (aged 1-24 months) seen in the childcare outpatient clinic of a pediatric hospital, from October to December of 2007. Using the same inhalation technique recommended for treatment of respiratory diseases and asthma in this age group, four holding chambers were analyzed: Flumax Baby, AlergoChamber, LuftChamber, and ACE (n = 20 infants per analysis group). The variables measured were gradient of inspiratory pressure and time for the valve to open. The holding chambers did not differ in gradient of inspiratory pressure (p = 0.382) or time to open the valve (p = 0.199). Fifteen of 80 infants were not able to generate sufficient inspiratory pressure to open the valve (1 using AlergoChamber, 5 using Flumax Baby, 6 using ACE, and 3 using LuftChamber), with no significant differences between devices. The holding chambers were similar regarding pressure gradient and time required to open the valve. The fact that 19% of the sample was unable to open the valve suggests that it would be useful to determine the ability of small children (<24 months of age) to use holding chambers before prescribing them.
机译:本研究旨在确定婴儿是否产生足够的吸气压力梯度来打开各种市售容纳室的单向阀,以及达到该梯度所需的时间。 2007年10月至2007年12月,在儿科医院的托儿门诊中对80名健康婴儿(1-24个月大)进行了一项前瞻性横断面研究。采用推荐的相同吸入技术治疗呼吸系统疾病和在这个年龄组的哮喘患者中,分析了四个容纳室:Flumax Baby,AlergoChamber,LuftChamber和ACE(每个分析组n = 20婴儿)。测量的变量是吸气压力的梯度和阀门打开的时间。储气室的吸气压力梯度(p = 0.382)或打开阀门的时间(p = 0.199)没有差异。 80个婴儿中有15个无法产生足够的吸气压力来打开瓣膜(1个使用AlergoChamber,5个使用Flumax Baby,6个使用ACE,3个使用LuftChamber),设备之间无显着差异。保持腔在压力梯度和打开阀所需的时间方面相似。 19%的样本无法打开阀门这一事实表明,在开处方之前,确定小孩(<24个月大)使用储藏室的能力将非常有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号