首页> 外文期刊>Journal of Zoo and Wildlife Medicine >COMPARISON OF A SMARTPHONE-BASED ELECTROCARDIOGRAM DEVICE WITH A STANDARD SIX-LEAD ELECTROCARDIOGRAM IN THE ATLANTIC BOTTLENOSE DOLPHIN ( TURSIOPS TRUNCATUS).
【24h】

COMPARISON OF A SMARTPHONE-BASED ELECTROCARDIOGRAM DEVICE WITH A STANDARD SIX-LEAD ELECTROCARDIOGRAM IN THE ATLANTIC BOTTLENOSE DOLPHIN ( TURSIOPS TRUNCATUS).

机译:基于智能手机的心电图装置比较了大西洋瓶装海豚(Tursiops Truncatus)的标准六引线心电图。

获取原文
获取原文并翻译 | 示例
           

摘要

A bipolar, single-lead electrocardiogram (ECG) device is capable of recording ECGs with an integrated smartphone application. To determine the utility of this device, phone-based ECGs (pECG) were compared with standard six-lead ECGs (sECG) in four female Atlantic bottlenose dolphins ( Tursiops truncatus) at the National Aquarium. Study animals were trained to haul out onto a dry deck in ventral recumbency and allow simultaneous 30-sec ECG acquisition using the two devices. The pECG device was held against the thoracic wall caudal to the left axilla. The sECGs were recorded in the frontal plane. Instantaneous heart rates were obtained from identical QRS complexes on both ECGs. Three boarded cardiologists independently evaluated the rhythm and the polarity of the QRS depolarization for each recording and the results were compared. The mean heart rate was 80 beats/min (range 62-92 beats/min) and 80 beats/min (range 60-92 beats/min) for the pECG and sECGs, respectively. All four dolphins displayed sinus respiratory arrhythmia, and one animal had occasional atrial premature contractions. Rhythm diagnosis and QRS polarity were identical for the pECG and sECG. Dolphin vocalizations created artifacts on the pECG that were not present on the sECG, so pECGs had to be acquired without vocalization. The pECG appears to be an accurate and useful method to monitor the heart rate in Atlantic bottlenose dolphins. This system is inexpensive and portable, making it valuable for health examinations, transport monitoring, and stranding responses.
机译:双极,单引灯心电图(ECG)设备能够使用集成的智能手机应用记录ECG。为了确定该设备的效用,将基于电话的ECG(PECG)与在国家水族馆的四个女性大西洋瓶颈海豚(Tursiops Truncatus)中的标准六个引导ECG(SECG)进行了比较。学习动物培训,训练到腹侧校正的干燥甲板上,并允许使用两种器件同时进行30秒的ECG采集。 PECG装置将胸壁尾部保持在左腋窝。 Secgs被记录在正面平面中。在两个心电片上从相同的QRS复合物获得瞬时心率。三位登机心理学家独立评估节奏和QRS对每个记录的QRS去极化的极性,并进行了结果。平均心率为80次/分钟(范围62-92次/分钟),分别为PECG和SECGS的80次/分钟(范围60-92次/分钟)。所有四只海豚均显示鼻窦呼吸道性心律失常,一只动物偶尔会出现心房过早收缩。节奏诊断和QRS极性对于PECG和SECG是相同的。海豚发声创建了在SECG上不存在的PECG上的文物,因此必须在没有发声的情况下获得PECG。 PECG似乎是监测大西洋瓶颈海豚的心率准确和有用的方法。该系统便宜且便携式便携,使健康检查,运输监测和搁浅的反应有价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号