首页> 外文期刊>Journal of Thermal Spray Technology >Analysis of Corrosion Behavior and Surface Properties of Plasma-Sprayed HA/Ta Coating on CoCr Alloy
【24h】

Analysis of Corrosion Behavior and Surface Properties of Plasma-Sprayed HA/Ta Coating on CoCr Alloy

机译:COOL合金腐蚀行为及腐蚀行为及表面性质的分析

获取原文
获取原文并翻译 | 示例
           

摘要

The cobalt-chromium (CoCr) alloys have been extensively used as implants, especially in total joint replacements and in odontology, due to their superior mechanical properties and wear resistance in vivo. However, the excessive release of the Co' and Cr' ions from CoCr implants can lead to adverse health issues, such as hypersensitivity and inflammatory reactions. The present study aimed to improve the corrosion resistance of a medical grade CoCr alloy (ASTM F-1537) plasma-sprayed with tantalum (Ta)-reinforced hydroxyapatite (HA) coating. The weight percent (wt.%) of Ta content in HA coating was varied at three levels, i.e., 10, 20, and 30%. In vitro corrosion behavior was investigated by electrochemical measurements in Ringer's solution along with surface properties analysis. The results revealed an increase in surface hardness value with an incremental increase in Ta content in the HA coating. The surface of HA as well as Ta-reinforced HA coatings possessed adequate roughness and demonstrated hydrophilic nature. With the Ta reinforcement in HA coating, the E-corr values shifted toward nobler potentials and I-corr values declined noticeably which indicated an increase in corrosion resistance of the surface. The results of the study indicate that the proposed Ta reinforcement in HA is potentially important for CoCr bio-implant applications.
机译:钴 - 铬(COCR)合金已被广泛地用作植入物,特别是在总关节置换中和神话中,由于它们的机械性能优越,耐磨性在体内。然而,Co'和Cr'离子的过度释放来自Cox植入物可以导致不良健康问题,例如过敏反应和炎症反应。本研究旨在提高医用级CoCrogog(ASTM F-1537)血浆喷涂钽(TA) - 粘胺(HA)涂层的耐腐蚀性。在HA涂层中的Ta含量的重量百分比(重量%)在三个水平下变化,即10,20和30%。通过Ringer溶液中的电化学测量以及表面特性分析来研究体外腐蚀行为。结果表明,表面硬度值的增加,HA涂层中的TA含量增加增加。 HA的表面以及Ta-钢筋HA涂层具有足够的粗糙度并证明了亲水性。通过在HA涂层中的TA加固,E-Corr值朝高高贵的电位和I-Corr值下降显着下降,表明表面的耐腐蚀性增加。该研究的结果表明,HA中所提出的TA加固可能对COCR生物植入应用可能是重要的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号