Adrenal Biopsy under Wide-Bore MR Imaging Guidance

机译:宽孔MR成像指导下的肾上腺活组织检查

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Twenty-four magnetic resonance (MR) imaging-guided percutaneous adrenal biopsies performed between April 2009 and October 2016 were reviewed retrospectively. Epidemiologic, procedural, and histopathologic data were retrospectively collected. Mean size of tumors was 4.3 cm (range, 1.5-16.0 cm). Mean procedure time was 49 min (range, 24-95 min). Mean needle angulation was 27.7 degrees (range, 0 degrees-60 degrees). Mean depth was 9.6 cm (range, 5.8-13.7 cm). The sensitivity, specificity, positive predictive value, negative predictive value, and accuracy of MR imaging-guided biopsy were 95.5%, 100%, 100%, 66.7%, and 95.8%, respectively. There were no immediate or delayed complications. MR imaging guidance seems safe and accurate to target adrenal-gland masses.
机译:回顾性地审查了2009年4月和2016年10月在2016年4月至2016年4月之间进行的二十四个磁共振(MR)的经皮肾上腺活检。 回顾性地收集流行病学,程序和组织病理数据。 肿瘤的平均尺寸为4.3厘米(范围,1.5-16.0cm)。 平均程序时间为49分钟(范围,24-95分钟)。 平均针角度为27.7度(范围,0度-60度)。 平均深度为9.6厘米(范围,5.8-13.7厘米)。 MR成像引导活检的敏感性,特异性,阳性预测值,阴性预测值和准确性分别为95.5%,100%,100%,66.7%和95.8%。 没有立即或延迟并发症。 MR成像指导似乎安全,准确地靶向肾上腺腺体。