首页> 外文期刊>Journal of the Optical Society of America, A. Optics, image science, and vision >Comparison of refractive error measurements by three different models of autorefractors and subjective refraction in young adults
【24h】

Comparison of refractive error measurements by three different models of autorefractors and subjective refraction in young adults

机译:三种不同型号的刷新误差测量与年轻成年人的主观折射的比较

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, the accuracy of noncycloplegic refractive error measurements performed by three autorefractors, Nidek ARK-510A, Visionix L80 Wave+, and Shin-Nippon NVision-K 5001, compared with subjective refraction (SR) in young adults is analyzed using statistical methods. The analysis is carried out in terms of spherical equivalent (SE), astigmatism, and two Jackson crossed-cylinder vectors. Fifty-one subjects aged 19-23 were examined. The differences between objective refraction (OR) measured by autorefractors are not statistically significant in either SE or cylindrical components. However, in comparison with subjective refraction, the best agreement in SE was provided by K 5001. For all instruments, differences in SE between OR and SR are statistically significant. (C) 2019 Optical Society of America
机译:本文采用统计方法分析了三种易用子,NIDEK ARK-510A,Visionix L80波+和Shin-Nippon NVision-K 5001进行的非狭窄术折射误差测量的准确性。 分析在球形等同物(SE),散光和两个杰克逊交叉缸向量方面进行。 检查了19-23岁的五十个科目。 通过自动除颤器测量的客观折射(或)之间的差异在SE或圆柱形部件中没有统计学意义。 然而,与主观折射相比,通过K 5001提供SE中的最佳协议。对于所有乐器,SE之间的差异在统计上有统计学意义。 (c)2019年光学学会