MAX IV is a fourth-generation, or diffraction-limited, synchrotron light source with a number of state-of-the-art beamlines. The performance of a beamline is, to a high degree, set by the energy resolution it can achieve, which in turn is governed to a large extent by the monochromator. During the design phase of a monochromator, the mechanical requirements must be fully understood and met with margin. During commissioning, the performance must be verified and optimized. In this paper, six soft X-ray monochromators at MAX IV beamlines (Bloch, Veritas, HIPPIE, SPECIES, FinEstBeAMS and SoftiMAX) are examined with a focus on their resolving power, energy range and the time required to change measurement range, as those parameters are dependent on each other. The monochromators have a modern commercial design, planned and developed in close collaboration with the vendors. This paper aims to present the current status of the commissioning at MAX IV with emphasis on elucidating the mechanical limitations on the performance of the monochromators. It contains analysis of the outcome and our approach to achieve fast and high-resolution monochromators.
机译:MAX IV是具有多个最先进的束线的第四代或衍射限制的同步光源。光束线的性能是通过能量分辨率设定的,它可以实现的能量分辨率,这反过来又在很大程度上被单色器控制。在单色仪的设计阶段,必须完全理解机械要求并满足边缘。在调试期间,必须验证和优化性能。在本文中,在MAX IV波束线(Bloch,Veritas,Hippie,物种,FinestBeams和Softimax)的六个软X射线单色器进行了检查,重点放在它们的分辨率,能量范围和更换测量范围所需的时间参数彼此依赖。单色器具有现代化的商业设计,计划和开发与供应商密切合作。本文旨在提出最大IV的调试现状,重点阐明了对单色仪性能的机械限制。它包含了实现快速和高分辨率单色器的结果的分析和我们的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号