In vitro evaluation of mechanical characteristics of a novel Patent Ductus Arteriosus (PDA) occlusion device with a new lateral pressure measurement study.

机译:新型横向压力测量研究的新型专利导管动脉术(PDA)闭塞装置的机械特性进行体外评价。

获取原文
获取原文并翻译 | 示例
           

摘要

Persistence of the ductus arteriosus (DA), connecting the aorta to the pulmonary artery, after birth leads to a congenital heart disease known as Patent Ductus Arteriosus (PDA). The objective of this study was to propose a new intraductal design for PDA occlusion and determine the efficacy of the new device by means of an in vitro mechanical evaluation. A novel intraductal design device called the Novel design 36 (ND36) was proposed, fabricated and tested. Mechanical tests included tests to determine the mechanical characteristics, force-displacement characteristics, stiffness, device slenderization, minimum diameter attained, insertion force required to pull device at site and lateral pressure exerted on the PDA walls. Comparative analysis was carried out with the Amplatzer Duct Occluder (AGA Medical Corporation, Golden Valley, MN). The results showed that the ND36 when fully stretched was shorter than the ADO implying an easier in vivo movement. The minimum diameter attained by ND36 is 2.52 mm which is compatible with standard catheter bores of 6-8 Fr (2-2.7 mm) bore sizes. A new setup was devised in order to carry out comparative lateral pressure measurements between the ND36 and ADO. The results showed that the ND36 exerted a pressure of 10 KPa (=75 mm Hg) and the ADO exerted 80 KPa (=600 mm Hg) over the walls of the tubes which is a physiologically promising result for the ND36 as it shows that the ND36 would exert lesser stress on the walls of the duct. The results suggest that the novel intraductal design (ND36) would prove to be a safer, reliable and effective device for PDA occlusion.
机译:在出生后将主动脉连接到肺动脉的导管血管(DA)持久性导致先天性心脏病,称为专利导管术(PDA)。本研究的目的是为PDA闭塞提出新的内部内部设计,并通过体外机械评估确定新设备的功效。提出了一种新颖的内部设计装置,称为新颖的设计36(ND36),制造和测试。机械测试包括确定机械特性,力 - 位移特性,刚度,装置细长,达到的最小直径,在PDA壁上施加的横向压力所需的插入力。比较分析与Amplatzer管道封堵器(AGA Medical Corporation,Golden Valley,Mn)进行。结果表明,当完全拉伸时,ND36短于ADO暗示体内运动更容易。 ND36获得的最小直径为2.52毫米,其与6-8FR(2-2.7mm)孔尺寸的标准导管孔兼容。设计了一个新的设置,以便在ND36和ADO之间进行比较横向压力测量。结果表明,ND36施加10kPa(= 75mm Hg)的压力,并且ADO在管的壁上施加80kPa(= 600mm Hg),这是ND36的生理上有希望的结果ND36将对管道的墙壁上发挥较小的压力。结果表明,新颖的内部内部设计(ND36)将被证明是PDA闭塞的更安全,可靠和有效的装置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号