首页> 外文期刊>Journal of orthopaedic trauma >Experimental study of distal interlocking of a solid tibial nail: radiation-independent distal aiming device (DAD) versus freehand technique (FHT).
【24h】

Experimental study of distal interlocking of a solid tibial nail: radiation-independent distal aiming device (DAD) versus freehand technique (FHT).

机译:固体胫骨钉远端互锁的实验研究:辐射无关的远端瞄准装置(爸爸)与手法技术(FHT)。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: Recently, radiation-independent aiming devices for the tibia that compensate for insertion-related implant deformation have been developed, but the benefits of such systems have not been determined. This study prospectively evaluated the duration of the nailing procedure, the length of radiation time, and the accuracy of interlocking screw placement when using a radiation-independent distal aiming device (DAD) versus the freehand technique (FHT). MATERIALS AND METHODS: In an oblique cadaveric tibial fracture, a surgeon inexperienced in both techniques (DAD and FHT) performed statically locked intramedullary nailing. RESULTS: For the DAD and the FHT, respectively, the total operation time was 25.4 +/- 11.3 (mean +/- standard deviation) versus 30.9 +/- 14.3 minutes (p = 0.029), the distal locking time was 16.7 +/- 8.6 versus 21.9 +/- 10.5 minutes (p = 0.004), the total fluoroscopy time was 9 +/- 5 versus 93 +/- 34 seconds (p < 0.0001), the distal locking fluoroscopy time was zero versus 88 +/- 33 seconds (p < 0.0001), and the screw wear was -0.7 +/- 5.2 versus 26.8 +/- 31.6 micrometers (p = 0.001). The failure rate was 1.6 percent (one of sixty screws) in both groups. CONCLUSION: These results suggest that the DAD can eliminate the need for radiation during placement of distal interlocking screws.
机译:目的:最近,已经开发出用于补偿插入相关的植入变形的胫骨的无关的瞄准装置,但这些系统的益处尚未确定。本研究预期评估了钉指过程的持续时间,辐射时间的长度,以及在使用辐射无关的远端瞄准装置(爸爸)与手法技术(FHT)时互锁螺钉放置的准确性。材料和方法:在倾斜的尸体胫骨骨折中,两种技术(爸爸和FHT)缺乏经验的外科医生在静态锁定的髓内钉上进行。结果:对于爸爸和FHT,总操作时间分别为25.4 +/- 11.3(平均+/-标准偏差)与30.9 +/- 14.3分钟(P = 0.029),远端锁定时间为16.7 + / - 8.6与21.9 +/-105分钟(p = 0.004),总荧光透视时间为9 +/- 5,与93 +/- 34秒(p <0.0001),远端锁定透视时间为零,而88 +/- 33秒(P <0.0001),螺杆磨损为-0.7 +/- 5.2与26.8 +/- 31.6微米(p = 0.001)。两组的故障率为1.6%(60个螺钉中的一个)。结论:这些结果表明,爸爸在放置远端互锁螺钉期间可以消除对辐射的需求。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号