Respiratory rate (RR) is an important vital sign used in the initial and ongoing assessment of acutely ill patients. It is also used as a predictor of serious deterioration in a patient's clinical condition. Convenient electronic devices exist for measurement of pulse, blood pressure, oxygen saturation, and temperature. Although devices which measure RR exist, none has entered everyday clinical practice. We have developed a contactless portable respiratory rate monitor (CPRM) and evaluated the agreement in respiratory rate measurements between existing methods and our new device. The CPRM uses thermal anemometry to measure breath signals during inspiration and expiration. RR data were collected from 52 healthy adult volunteers using respiratory inductance plethysmography (RIP) bands (established contact method), visual counting of chest movements (established noncontact method), and the CPRM (new method), simultaneously. Two differently shaped funnel attachments to the CPRM were evaluated for each volunteer. Data showed a good agreement between measurements from the CPRM and the gold standard RIP, with intraclass correlation coefficient (ICC): 0.836, mean difference 0.46 and 95% limits of agreement of -5.90 to 6.83. When separate air inlet funnels of the CPRM were analyzed, stronger agreement was seen with an elliptical air inlet; ICC 0.908, mean difference 0.37 with 95% limits of agreement -4.35 to 5.08. A contactless device for accurately and quickly measuring respiratory rate will be an important triage tool in the clinical assessment of patients. More testing is needed to explore the reasons for outlying measurements and to evaluate in the clinical setting.
机译:呼吸率(RR)是在急性病患者的最初和持续评估中使用的重要生命身材。它也被用作患者临床状况严重恶化的预测因子。可以方便的电子设备用于测量脉冲,血压,氧气饱和度和温度。虽然存在测量RR的设备,但没有人进入日常临床实践。我们开发了非接触式便携式呼吸率监测率监测率监测率(CPRM),并在现有方法和新设备之间评估呼吸速率测量协议。 CPRM使用热风测定法测量灵感和呼气期间的呼吸信号。使用呼吸电感过度体积(RIP)带(已建立的接触方法),从52个健康成人志愿者(建立的接触方法),胸部运动(建立的非接触式方法)和CPRM(新方法),同时从52个健康的成人志愿者收集了RR数据。对CPRM的两个不同形状的漏斗附件被评估为每个志愿者。数据在CPRM和黄金标准RIP的测量之间显示了良好的一致性,其中包括脑内相关系数(ICC):0.836,平均差异0.46和95%的协议限值-5.90至6.83。当分析CPRM的单独进气漏斗时,用椭圆空气入口看到更强的一致性; ICC 0.908,平均差异0.37,协议的95%限制-4.35至5.08。用于准确且快速测量呼吸速率的非接触式设备将是患者临床评估中的重要分类工具。需要更多的测试来探讨广阔的测量和在临床环境中评估的原因。

著录项

  • 来源
    《Journal of Medical Devices》 |2020年第3期|共6页
  • 作者单位

    Respiratory Unit Sheffield Children's NHS Foundation Trust;

    Respiratory Unit Sheffield Children's NHS Foundation Trust;

    Industry and Innovation Research Institute Sheffield Hallam University;

    Emergency Department Sheffield Children's NHS Foundation Trust;

    Research and Innovation Office Cranfield University;

    Department of Engineering and Mathematics Sheffield Hallam University;

    Department of Materials and Engineering Research Institute Sheffield Hallam University;

    Design Futures Sheffield Hallam University;

    Department of Mechanical and Aerospace Engineering University of Strathclyde;

    Respiratory Unit Sheffield Children's NHS Foundation Trust;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 医用一般科学;
  • 关键词

  • 入库时间 2022-08-20 09:33:25

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号