首页> 外文期刊>Journal of instrumentation: an IOP and SISSA journal >Demonstration of an absolute Rayleigh scattering spectral calibration on the W7-X Thomson scattering system
【24h】

Demonstration of an absolute Rayleigh scattering spectral calibration on the W7-X Thomson scattering system

机译:W7-X Thomson散射系统上的绝对瑞利散射光谱校准的演示

获取原文
获取原文并翻译 | 示例
           

摘要

A novel in-situ calibration method for Thomson scattering (TS) diagnostics has been demonstrated on theWendelstein 7-X stellarator (W7-X). Utilizing a laser-pumped optical parametric oscillator (OPO), Rayleigh scattering measurements were performed in-situ in argon, nitrogen, and air at pressures from 0 to 1024 mbar and wavelengths between 730 and 1070 nm. As the OPO follows the same beam path as the TS lasers, the scattering angle and solid angle of collection can be made identical to the TS geometry; therefore, the calibration covers all optical components within the TS system, including the first vacuum window, and can, in principle, be used to directly calibrate each filter channel for absolute plasma density. Despite high background stray light levels, the Rayleigh scattered signal was successfully isolated from the stray light, and filter curves for multiple polychromators were resolved. The complete calibration system, planned for future W7-X operational campaigns, will aim to further reduce stray light, thereby decreasing the required gas pressure and allowing in-situ calibration measurements on service days. This is planned in part by better matching the OPO beam alignment, size, and divergence to the installed Nd:YAG TS lasers. In addition, the OPO bandwidth will be improved by using an injection seeder to increase filter edge resolution.
机译:关于汤森散射(TS)诊断的新型原位校准方法已经在恒温斯坦7-X螺旋螺栓(W7-X)上进行了说明。利用激光泵浦光学参数振荡器(OPO),在氩气,氮气和空气中以0至1024毫巴的压力和730和1070nm之间的波长在原位进行瑞利散射测量。当OPO遵循与TS激光器相同的光束路径时,可以与TS几何形状相同的散射角和固体角度;因此,校准覆盖TS系统内的所有光学组件,包括第一真空窗口,并且原则上可以用于直接校准每个过滤通道以进行绝对等离子体密度。尽管背景杂散光线水平高,但从杂散光成功隔离瑞利散射信号,并解决多种多色器的过滤曲线。计划用于将来W7-X运营活动计划的完整校准系统将旨在进一步减少杂散光,从而降低所需的气体压力并在服务日内允许原位校准测量。这部分通过更好地匹配OPO光束对准,尺寸和发散到安装的ND:YAG TS激光器的更好的计划。此外,通过使用喷射播种机可以提高opo带宽以增加滤波器边缘分辨率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号