首页> 外文期刊>Journal of endovascular therapy: an official journal of the International Society of Endovascular Specialists >Air Embolism During TEVAR: Liquid Perfluorocarbon Absorbs Carbon Dioxide in a Combined Flushing Technique and Decreases the Amount of Gas Released From Thoracic Stent-Grafts During Deployment in an Experimental Setting
【24h】

Air Embolism During TEVAR: Liquid Perfluorocarbon Absorbs Carbon Dioxide in a Combined Flushing Technique and Decreases the Amount of Gas Released From Thoracic Stent-Grafts During Deployment in an Experimental Setting

机译:Tevar期间的空气栓塞:液体全氟化碳在组合的冲洗技术中吸收二氧化碳,并在实验环境中展开期间从胸前移植物中释放的气体量降低

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose:To investigate the influence of flushing thoracic stent-grafts with carbon dioxide and perfluorocarbon on the amount of gas released during stent-graft deployment in thoracic endovascular aortic repair (TEVAR). Materials and Methods: Ten TX2 ProForm thoracic stent-grafts were deployed into a water-filled container with a curved plastic pipe and flushed sequentially with carbon dioxide, 20 mL of liquid perfluorocarbon (PFC), and 60 mL of saline. Released gas was measured using a calibrated setup. The volume of released gas was compared with the results of an earlier published reference group, in which identical stent-grafts were flushed with 60 mL saline alone as recommended in the instructions for use. Results: The average amount of gas released in the test group was 0.076 mL, significantly lower (p0.001) than the mean 0.79 mL of gas released in the reference group. Big bubbles appearing at the tip of the sheath when deployment was started were seen in all grafts of the reference group but in only 2 of the test group stent-grafts. Small bubbles were less frequent in the test group. Conclusion: The amount of gas released from thoracic stent-grafts during deployment can be influenced by different flushing techniques. The use of PFC in addition to the carbon dioxide flushing technique reduces the volume of gas released during deployment of tubular thoracic stent-grafts to a few microliters. This significant effect is presumably based on the high solubility of carbon dioxide in perfluorocarbon and could be a potential future approach to lower the risk of cerebral injury and stroke from air embolism during TEVAR.
机译:目的:探讨冲洗胸间支架与二氧化碳和全氟化碳的影响对胸腔血管内主动脉修复(TEVAR)支架施工期间释放的气体量。材料和方法:将十TX2骨压胸支架移植物展开到具有弯曲塑料管的充水容器中,并用二氧化碳依次冲洗,20ml液体全氟碳(PFC)和60毫升盐水。使用校准设置测量释放的气体。将释放的气体的体积与早期公开的参考组的结果进行比较,其中在使用说明书中,单独用60ml盐水冲洗相同的支架移植物。结果:试验组中释放的平均气体量为0.076mL,显着降低(P <0.001),比参考组中释放的平均气体。在参考组的所有移植物中可以看到鞘尖时出现在护套尖端的大气泡,但只有2个试验组支架移植物。在试验组中小气泡频率较小。结论:在部署期间从胸支架移植物释放的气体量可以受到不同冲洗技术的影响。除了二氧化碳冲洗技术之外,PFC除了二氧化碳冲洗技术之外还减少了在部署管状胸间支架移植到几个微升过程中释放的气体的体积。这一显着效果是基于全氟化碳中二氧化碳的高溶解度,并且可能是降低Tevar期间从空气栓塞的脑损伤和中风风险的潜在未来的方法。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号