PURPOSE: To present a modification of both the guiding catheter and stent delivery system to facilitate access into sharply curved supra-aortic vessels during carotid angioplasty. TECHNIQUE: Access failures during carotid interventions typically occur because either the vessel origin is acutely angled or the stenosis is located proximal to the carotid bifurcation, limiting access to the external carotid artery. We directly catheterized the left common carotid artery or brachiocephalic trunk with a guiding catheter whose tip had been curved by steam-shaping. The tip of the stent delivery system also was molded into a curve by this process so it engaged the curved part of the guiding catheter without pulling it out of the vessel. CONCLUSIONS: This simple modification can improve the success rate of carotid cannulation via the femoral approach and does not seem to increase the risk of the intervention.
机译:目的:呈现引导导管和支架递送系统的修改,以便于颈动脉血管成形术期间进入急剧弯曲的同学血管。 技术:颈动脉干预期间的进入故障通常发生,因为血管来源急剧成角度或狭窄位于颈动脉分叉的近端,限制了外部颈动脉。 我们直接导管左侧常见的颈动脉或肱骨牙齿行李箱,引导导管,其尖端通过蒸汽成形弯曲。 支架递送系统的尖端也通过该过程模制成曲线,使其接合引导导管的弯曲部分而不将其从容器中拉出。 结论:这种简单的改性可以通过股骨方法提高颈动脉插管的成功率,似乎不会增加干预的风险。