Purpose: To evaluate the efficacy of a new guidewire threading instrument (GTI) in reducing guidewire threading times compared with the traditional freehand method. Materials and Methods: This prospective, controlled, single-center study recruited 100 subjects (mean age 45.2 +/- 11.3 years; 43 men) and divided them into 2 groups depending on their professional background: 50 experienced medical personnel (performed or assisted in >50 catheter-based procedures) and 50 medical personnel with no experience in endovascular techniques (inexperienced group). The threading time of both groups was recorded using the conventional freehand method and with the GTI for both 0.035- and 0.014-inch platforms. Users of eyeglasses for presbyopia were tested with and without glasses. Median values are reported with the interquartile range (IQR; Q1, Q3) in parentheses. Results: The mean insertion times with both the 0.035-inch and 0.014-inch guidewires in the overall study group were significantly better with the use of the GTI compared with the traditional freehand method (p<0.001). Both the experienced and the inexperienced participants improved their insertion times using the GTI with both guidewire platforms (p<0.001). The threading time with the new device was also significantly reduced (p<0.001) for both participants with presbyopia and those without. When comparing the median absolute time improvement (difference between freehand/GTI insertion times) for the 0.014-inch platform, the inexperienced group showed a greater improvement in their performance [3.52 seconds (IQR 2.76, 5.12)] compared with the experienced group [1.87 seconds (IQR 1.37, 2.66), p<0.001]. The median absolute time improvement was also significantly greater for the presbyopic group [5.75 seconds (IQR 3.14, 8.20)] vs the group without age-related visual impairment [2.64 seconds (IQR 1.65, 3.36), p<0.001]. Conclusion: This simple and inexpensive homemade device facilitates wire threading of low-profile catheters and seems to be especially helpful for trainees with no experience and presbyopic operators.
机译:目的:与传统的手绘方法相比,评估新导丝穿线仪(GTI)减少导丝穿线时间的功效。材料和方法:这项前瞻性,控制,单中心研究招募了100名科目(平均年龄45.2 +/- 11.3岁; 43名男子),并根据他们的专业背景分为2组:50名经验丰富的医务人员(执行或协助> 50个基于导管的程序)和50名医务人员,内血管技术没有经验(缺乏经验的组)。使用传统的手法方法和GTI进行0.035和0.014英寸平台的GTI记录两组的穿线时间。用眼镜测试老年眼镜的用户。报告中位数值在括号中使用胎面范围(IQR; Q1,Q3)。结果:与传统的手法法相比,使用GTI(P <0.001)使用GTI,在整个研究组中的平均插入时间与总研究组的0.35英寸和0.014英寸导丝显着更好(P <0.001)。经验丰富的和缺乏经验的参与者使用GTI与导游平台(P <0.001)改进了它们的插入时间。对于具有老花​​眼的参与者以及没有的参与者,对新设备的穿线时间也显着降低(P <0.001)。在比较0.014英寸平台的中值绝对时间改进(FreeHand / GTI插入时间之间的差异)时,不经验的组在其性能方面表现出更大的改善[3.52秒(IQR 2.76,5.12)]与经验组[1.87秒(IQR 1.37,2.66),P <0.001]。高调组的中位绝对时间改善也明显更大[5.75秒(IQR 3.14,8.20)]与没有年龄相关的视觉损伤的组[2.64秒(IQR 1.65,3.36),P <0.001]。结论:这种简单廉价的自制装置有助于低调导管的丝线,似乎对没有经验和老板运营商的学员特别有帮助。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号