首页> 外文期刊>Journal of chromatography, B. Analytical technologies in the biomedical and life sciences >Simple geometrical modifications for substantial color intensity and detection limit enhancements in lateral-flow immunochromatographic assays
【24h】

Simple geometrical modifications for substantial color intensity and detection limit enhancements in lateral-flow immunochromatographic assays

机译:侧流免疫色谱分析中大量色彩强度和检测限增强的简单几何修改

获取原文
获取原文并翻译 | 示例

摘要

One of the ongoing challenges in lateral flow Immunochromatographic assays (LFIA), is lowering the limit of detection and enhancing their signal quality, i.e.the color intensity.There are a number of rather costly and complicated processes for this aim, such as the use of functionalized materials/membranes and additional spectroscopic readout units.Nonetheless, there are simple and easy to practice alternatives, to be uncovered by analyzing the essential parameters of immunological reactions.The color intensity of the test line is a function of analytes flow velocity and their reaction rate.Detection pad width and test line position impact the flow velocity and reaction rate kinetics, examined in this paper for the limit of detection (LOD) and test-line color intensity.Firstly, the impact of width on the LOD was examined for human chorionic gonadotropin (pregnancy biomarker).Test line color intensity was measured using five different widths of the detection pad (trapezoidal) and four different test line positions, and the trends observed were explained according to the measured evolution of the velocity along the chromatography paper.With a constant width absorbent pad, LOD was cut by half to 5 mIU/ml by using a narrowing width detection pad, which keeps the wicking velocity higher than normal strips, and compared to them, color intensity increase between 55 and 150%, depending on the concentration.Nevertheless, a widening detection pad might cut the color intensity up to 150%, compared to normal strips, due to a profound decline of the analyte to ligand ratio at the test line.In addition, adequately sending the test line away from the conjugate pad yields the highest possible color intensity, for up to 400% of increase, in lower concentrations and narrowing test pads.However, further distancing the test line downfalls the color intensity.
机译:横向流动免疫层析测定(LFIa)中的持续挑战之一,降低了检测极限并提高了信号质量,Iethe颜色强度。此目的是许多相当昂贵和复杂的过程,例如使用官能化材料/膜和额外的光谱读出单位。无论如何,通过分析免疫反应的基本参数,尚未揭示替代方案。试线的颜色强度是分析物流速及其反应的函数速率。垫宽度和测试线位置会影响流速和反应速率动力学,在本文中检查了检测(LOD)和测试线颜色强度的限制。过度地,对人类的宽度对LOD的影响被用于人类使用五种不同宽度的检测垫(梯形)和四个D测量绒毛膜促性腺激素(妊娠生物标志物)。最终线颜色强度。根据沿色谱纸的速度的测量的速度解释,观察到的趋势。通过恒定的宽度吸收垫,通过使用狭窄的宽度检测垫将LOD切割成一半至5 miu / ml,将芯筒速度高于正常条带,并与它们相比,与浓度相比,颜色强度在55和150%之间增加,根据浓度,与正常条带相比,加宽检测垫可能会使彩色强度降低150%,因此在测试线上的分析物与配体比的深刻衰退。另外,将测试线充分发送远离共轭垫,产生最高可能的颜色强度,高达400%的增加,较低浓度和变窄的测试垫增加但是,如果以外,进一步疏远了测试线倒塌的颜色强度。

著录项