Vector analysis of astigmatic changes and optical quality outcomes after wavefront-guided laser in?situ keratomileusis using a high-resolution aberrometer

机译:波前引导激光器中的散光变化和光学质量结果的矢量分析?使用高分辨率异常量子测定的原位角膜瘤

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose To assess the ocular optical quality results as well as the?astigmatic changes after wavefront-guided laser in?situ keratomileusis (LASIK) in eyes with myopic astigmatism. Setting Instituto Clínico-Quirúrgico Oftalmológico, Bilbao, Vizcaya, Spain. Design Prospective case series. Methods Thirty-three eyes having wavefront-guided LASIK with the Star S4IR excimer laser combined with the iDesign system were enrolled in the study. Changes in visual acuity, refraction, higher-order aberrations, and ocular scattering index (OSI) were evaluated during a 3-month follow-up. The Alpins method was used to analyze the astigmatic changes. Results A significant reduction was observed in refraction ( P ? P ? P ?≥?.551). Likewise, no significant changes were observed in OSI ( P ?=?.361), with a mean 90-day postoperative value of 0.66?±?0.58 (SD). A significant change to negative values of the magnitude of error ( P ?=?.007) and a significant decrease in the correction index ( P ?=?.004) were observed during the follow-up, with mean 90-day postoperative values of ?0.24?±?0.28 D and 0.86?±?0.17 D, respectively. Conclusion Wavefront-guided LASIK using a high-resolution aberrometer provided safe and efficacious correction of myopic astigmatism, with preservation of the ocular optical quality.
机译:目的是评估眼睛光学质量的结果以及波前引导激光后的散差变化?在眼中的眼睛中的眼睛中的眼睛肌肉(LASIK)。设置Institutoclínico-quirúrgicooftalmológico,毕尔巴鄂,西班牙Vizcaya。设计前瞻性案例系列。方法采用波前引导LASIK的三十三只眼睛与idesign系统结合的明星S4IR准分子激光器进行了研究。在3个月的随访期间评估视力,折射,高阶像差和眼散射指数(OSI)的变化。 ALPINS方法用于分析散光变化。结果在折射中观察到显着的减少(p?p≤p≤≤551)。同样,在OSI(P?= 361)中没有观察到显着变化,平均90天术后值0.66?±0.58(SD)。在随访期间观察到误差幅度(P?=Δ.007)的负值的显着变化(P?=Δ.007)和校正指数(p?=Δ.004)的显着降低,平均90天术后值α0.24?±0.28 d和0.86?±0.17 d。结论使用高分辨率异室的波前引导LASIK提供了近视散光的安全和有效的矫正,并保存了眼部光学质量。

著录项

  • 来源
  • 作者单位

    From the Instituto Clínico-Quirúrgico Oftalmológico (Durán Gutiérrez Atienza) Bilbao the Basque;

    From the Instituto Clínico-Quirúrgico Oftalmológico (Durán Gutiérrez Atienza) Bilbao the Basque;

    From the Instituto Clínico-Quirúrgico Oftalmológico (Durán Gutiérrez Atienza) Bilbao the Basque;

    From the Instituto Clínico-Quirúrgico Oftalmológico (Durán Gutiérrez Atienza) Bilbao the Basque;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 眼科学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号