Field Detection of Drugs of Abuse in Oral Fluid Using the Alere (TM) DDS (R) 2 Mobile Test System with Confirmation by Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)

机译:使用Slere(TM)DDS(R)2移动测试系统在口服液中滥用滥用药物的田间检测,通过液相色谱串联质谱(LC-MS / MS)确认

获取原文
获取原文并翻译 | 示例
           

摘要

The collection and analysis of drugs in oral fluid (OF) at the roadside has become more feasible with the introduction of portable testing devices such as the Alere (TM) DDS (R) 2 Mobile Test System (DDS (R) 2). The objective of this study was to compare the on-site results for the DDS (R) 2 to laboratory-based confirmatory assays with respect to detection of drugs of abuse in human subjects. As part of a larger Institutional Review Board approved study, two OF samples were collected from each participant at a music festival in Miami, FL, USA. One OF sample was field screened using the DDS (R) 2, and a confirmatory OF sample was collected using the Quantisal (TM) OF collection device and submitted to the laboratory for testing. In total, 124 subjects participated in this study providing two contemporaneous OF samples. DDS (R) 2 field screening yielded positive results for delta-9-tetrahydrocannabinol (THC) (n = 27), cocaine (n = 12), amphetamine (n = 3), methamphetamine (n = 3) and benzodiazepine (n = 1). No opiate-positive OF samples were detected. For cocaine, amphetamine, methamphetamine and benzodiazepines, the DDS (R) 2 displayed sensitivity, specificity and accuracy of 100%. For THC, the DDS (R) 2 displayed sensitivity of 90%, specificity of 100% and accuracy of 97.5%, when the threshold for confirmation matched that of the manufacturers advertised cut-off. When this confirmatory threshold was lowered to the analytical limit of detection (i.e., 1 ng/mL), apparent device performance for THC was poorer due to additional samples testing positive by confirmatory assay that had tested negative on the DDS (R) 2, demonstrating a need for correlation between manufacturer cut-off and analytical reporting limit. These results from drug-using subjects demonstrate the value of field-based OF testing, and illustrate the significance of selecting an appropriate confirmation cut-off concentration with respect to performance evaluation and detection of drug use.
机译:在路边的口腔液中药物的收集和分析与引入便携式测试装置(如Salere(TM)DDS(R)2移动测试系统(DDS2))变得更加可行。本研究的目的是将DDS(R)2的现场结果与基于实验室的确认测定进行比较,检测人受试者中的滥用药物。作为较大的制度审查委员会批准的研究的一部分,来自美国迈阿密的音乐节的每个参与者收集了两种样品。使用DDS(R)2筛选样品之一,并且使用收集装置的量子(TM)收集样品的确认,并提交给实验室进行测试。总共参加了124名科目,参加了这项研究,提供了两种同时的样品。 DDS(R)2场筛选产生δ-9-四氢甘油(THC)(N = 27),可卡因(N = 12),氨基(N = 3),甲基苯丙胺(n = 3)和苯并二氮杂卓(n = 1)。检测到样品的呈态阳性。对于可卡因,安非他明,甲基苯丙胺和苯二氮噻嗪,DDS(R)2显示敏感性,特异性和准确性为100%。对于THC,DDS(R)2显示的灵敏度为90%,特异性100%,准确度为97.5%,当确认的阈值与制造商宣传截止的阈值相匹配时。当该确认阈值降低到检测的分析极限(即1ng / ml)时,由于在DDS(R)2上测试负数的额外样品,所以表现为THC的表观器件性能较差,这是在DDS(R)2上的负阴性,展示需要制造商切断与分析报告限制之间的相关性。来自药物的受试者的这些结果证明了基于现场的测试的价值,并说明了选择适当的确认截止浓度相对于性能评估和检测药物使用的重要性。

著录项

  • 来源
    《Journal of Analytical Toxicology》 |2018年第3期|共7页
  • 作者单位

    Fredr Rieders Family Fdn Ctr Forens Sci Res &

    Educ 2300 Stratford Ave Willow Grove PA 19090 USA;

    Fredr Rieders Family Fdn Ctr Forens Sci Res &

    Educ 2300 Stratford Ave Willow Grove PA 19090 USA;

    Fredr Rieders Family Fdn Ctr Forens Sci Res &

    Educ 2300 Stratford Ave Willow Grove PA 19090 USA;

    Fredr Rieders Family Fdn Ctr Forens Sci Res &

    Educ 2300 Stratford Ave Willow Grove PA 19090 USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 毒物学(毒理学);
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号