Effect of varying skin surface electrode position on electroretinogram responses recorded using a handheld stimulating and recording system

机译:使用手持式刺激和记录系统记录的不同皮肤表面电极位置对电气图响应的影响

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose A handheld device (the RETeval system, LKC Technologies) aims to increase the ease of electroretinogram (ERG) recording by using specially designed skin electrodes, rather than corneal electrodes. We explored effects of electrode position on response parameters recorded using this device. Methods Healthy adult twins were recruited from the TwinsUK cohort and underwent recording of light-adapted flicker ERGs (corresponding to international standard stimuli). In Group 1, skin electrodes were placed in a “comfortable” position, which was up to 20?mm below the lid margin. For subsequent participants (Group 2), the electrode was positioned 2?mm from the lid margin as recommended by the manufacturer. Amplitudes and peak times (averaged from both eyes) were compared between groups after age-matching and inclusion of only one twin per pair. Light-adapted flicker and flash ERGs were recorded for an additional 10 healthy subjects in two consecutive recording sessions: in the test eye, electrode position was varied from 2 to 10–20?mm below the lid margin between sessions; in the fellow (control) eye, the electrode was 2?mm below the lid margin throughout. Amplitudes and peak times (test eye normalised to control eye) were compared for the two sessions. Results Including one twin per pair, and age-matching yielded 28 individuals per group. Flicker ERG amplitudes were significantly lower for Group 1 than Group 2 participants ( p? =?0.0024). However, mean peak times did not differ between groups ( p? =?0.54). For the subjects in whom electrode position was changed between recording sessions, flash and flicker amplitudes were significantly lower when positioned further from the lid margin ( p? ?0.5). Conclusions Moving the skin electrodes further from the lid margin significantly reduces response amplitudes, highlighting the importance of consistent electrode positioning. However, this does not significantly affect peak times. Thus, it may be feasible to adopt a more comfortable position in participants who cannot tolerate the recommended position if analysis is restricted to peak time parameters.
机译:目的,手持设备(Reteval System,LKC Technologies)旨在通过使用专用设计的皮肤电极而不是角膜电极来提高ElectroreTino(ERG)记录的容易性。我们探讨了电极位置对使用该设备记录的响应参数的影响。方法采用双胞胎群招募健康的成人双胞胎,并经历了轻质闪烁ERG的记录(对应于国际标准刺激)。在第1组中,将皮肤电极放置在“舒适”位置,其高达盖子边缘低20Ωmm。对于随后的参与者(第2组),根据制造商推荐,电极从盖子边缘定位2Ωmm。比较在年龄匹配和每对中仅含一个双胞胎后的组之间比较幅度和峰时(从两只眼睛的平均)。在两个连续的记录会话中为另外的10个健康受试者记录了光调整的闪烁和闪存ERG:在测试眼中,电极位置在会话之间的盖子边缘下方的2至10-20毫米。在同伴(控制)的眼睛中,整个盖子边缘低2毫米。比较两个会话比较了两个会话的巨大和峰值时间(标准化以控制眼睛的试眼)。结果包括每对一对双胞胎,和年龄匹配的每组28个个体。对于第1组参与者,Flicker ERG振幅显着降低(P?= 0.0024)。然而,组的平均峰时间在组之间没有差异(p?= 0.54)。对于在记录会话之间改变电极位置的受试者,当进一步从盖子边缘进行定位时,闪光和闪烁幅度显着降低(P?0.5)。结论从盖子边缘进一步移动皮肤电极显着降低了响应幅度,突出了一致电极定位的重要性。但是,这不会显着影响高峰时间。因此,如果分析限于峰值时间参数,则在不能容忍推荐位置的参与者中采用更舒适的位置可能是可行的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号