【24h】

Integrated circuit detector technology in abdominal CT: Added value in obese patients

机译:腹部CT的集成电路检测器技术:肥胖患者的增值

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE. The purpose of this article was to assess the effect of an integrated circuit (IC) detector for abdominal CT on image quality. MATERIALS AND METHODS. In the first study part, an abdominal phantom was scanned with various extension rings using a CT scanner equipped with a conventional discrete circuit (DC) detector and on the same scanner with an IC detector (120 kVp, 150 effective mAs, and 75 effective mAs). In the second study part, 20 patients were included who underwent abdominal CT both with the IC detector and previously at similar protocol parameters (120 kVp tube current-time product and 150 reference mAs using automated tube current modulation) with the DC detector. Images were reconstructed with filtered back projection. RESULTS. Image quality in the phantom was higher for images acquired with the IC compared with the DC detector. There was a gradually increasing noise reduction with increasing phantom sizes, with the highest (37% in the largest phantom) at 75 effective mAs ( p < 0.001). In patients, noise was overall significantly ( p = 0.025) reduced by 6.4% using the IC detector. Similar to the phantom, there was a gradual increase in noise reduction to 7.9% in patients with a body mass index of 25 kg/m 2 or lower ( p = 0.008). Significant correlation was found in patients between noise and abdominal diameter in DC detector images (r = 0.604, p = 0.005), whereas no such correlation was found for the IC detector (r = 0.427, p = 0.060). CONCLUSION. Use of an IC detector in abdominal CT improves image quality and reduces image noise, particularly in overweight and obese patients. This noise reduction has the potential for dose reduction in abdominal CT.
机译:目的。本文的目的是评估用于腹部CT的集成电路(IC)检测器对图像质量的影响。材料和方法。在第一个研究部分中,使用配备了常规分立电路(DC)检测器的CT扫描仪和带有IC检测器(120 kVp,150有效mAs和75有效mAs)的同一扫描仪,用各种延长环扫描腹部体模)。在第二部分研究中,包括20名患者,他们都使用IC检测器进行腹部CT检查,并且先前都使用DC检测器以相似的协议参数(120 kVp电子管电流-时间乘积和150参考mAs使用自动电子管电流调制)进行了腹部CT检查。用滤波后的反投影重建图像。结果。与DC检测器相比,用IC采集的图像的幻像中的图像质量更高。随着幻像尺寸的增加,降噪逐渐增加,在75 mA有效电流下最大(最大幻像中为37%)(p <0.001)。在患者中,使用IC检测器可将噪音总体降低(p = 0.025)6.4%。与体模相似,体重指数在25 kg / m 2或更低的患者中,降噪效果逐渐提高至7.9%(p = 0.008)。在直流检测器图像中,患者的噪声与腹部直径之间存在显着相关性(r = 0.604,p = 0.005),而对于IC检测器,则没有发现这种相关性(r = 0.427,p = 0.060)。结论。在腹部CT中使用IC检测器可以改善图像质量并降低图像噪声,特别是对于超重和肥胖的患者。这种降噪可能会降低腹部CT的剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号