Determination of Examination-Specific Diagnostic Reference Level in Computed Tomography by A New Quality Control-Based Dose Survey Method

机译:一种新的基于质量控制的剂量测量方法测定计算断层扫描中的检查特异性诊断参考水平

获取原文
获取原文并翻译 | 示例

摘要

A new quality-control-based (QC-based) dose survey method has been developed for determination of diagnostic reference levels (DRL) in Computed Tomography (CT) examinations. The QC-based dose survey method is based on the use of retrospective data in the QC documents and reports, which are typically available from the National Regulatory Authority database. The method was applied to 70 CT scanners in Tehran, Iran, by using the available QC reports from the database. The commonly used data collection method was also applied by filling each questionnaire on-site to validate the new method. Using the new QC-based and data collection methods, the DRLs of four common CT examinations: head, sinus, chest, and abdomen/pelvis were determined and compared. The DRLs determined by the QC-based method for head, sinus, chest, and abdomen/pelvis are 59, 29, 10, and 13 mGy, respectively, for the volume computed tomography dose index (CTDIVol) and 834, 235, 233, and 522 mGy-cm for the dose length product (DLP), respectively. The difference between the DRLs obtained by the two methods is on the average 6.7 +/- 5.7%, which is within the acceptance tolerance level of the IAEA for QC dosimetry tests. The QC-based dose survey method is believed to be an effective alternative method to the other commonly used data collection and direct dose measurement method for determination of CT examination DRLs. This new method has unique characteristics such as simplicity, time and cost effectiveness, highly reduced clinical interruptions and collaborations, and potential for large-scale surveys with capability for more frequent review of national DRL values.
机译:已经开发了一种基于质量控制的(基于QC的)剂量测量方法,用于测定计算断层扫描(CT)检查中的诊断参考水平(DRL)。基于QC的剂量调查方法基于QC文件和报告中的回顾数据的使用,通常可从国家监管机构数据库中获得。通过使用来自数据库的可用QC报告,将该方法应用于伊朗的德黑兰,伊朗的70 CT扫描仪。通过填充每个问卷的现场验证新方法,还应用了常用的数据收集方法。使用新的基于QC和数据收集方法,确定了四种常见CT检查的DRL:头部,鼻窦,胸部和腹部/骨盆,并比较。由基于QC的头部,窦,胸部和腹部/骨盆/骨盆测定的DRL分别为59,29,10和13 MGY,分别用于体积计算断层摄影剂量指数(CTDIVOL)和834,235,233,为剂量长度产品(DLP)分别为522毫米-cm。通过两种方法获得的DRL之间的差异平均为6.7 +/- 5.7%,其在IAEA的接受耐受性水平范围内的QC剂量测定试验。据信基于QC的剂量测量方法是其他常用的数据收集和直接剂量测量方法的有效替代方法,用于测定CT检查DRL。这种新方法具有独特的特性,例如简单,时间和成本效益,高度降低的临床中断和合作,以及大规模调查的潜力,具有对国家DRL值更频繁地审查的能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号