首页> 外文期刊>AJR: American Journal of Roentgenology : Including Diagnostic Radiology, Radiation Oncology, Nuclear Medicine, Ultrasonography and Related Basic Sciences >Contrast media-doped hydrodissection during thermal ablation: Optimizing contrast media concentration for improved visibility on CT images
【24h】

Contrast media-doped hydrodissection during thermal ablation: Optimizing contrast media concentration for improved visibility on CT images

机译:消融过程中造影剂掺杂的水解剖:优化造影剂浓度以提高CT图像的可见度

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVE. The purpose of this study is to determine a concentration of iodinated contrast media in saline and 5% dextrose in water (D5W) for organ hydrodissection, a technique used to physically separate and protect tissues adjacent to thermal ablations. MATERIALS AND METHODS. A total of 28 samples were prepared from 1:1000-1:1 iohexol or iothalamate meglumine contrast media in either normal saline or D5W. Samples alone or juxtaposed with a homogeneous liver-mimicking phantom were imaged by CT using 80-120 kVp and 10-300 mAs. Mean CT numbers and noise were measured from the fluid, background air, phantom adjacent to the fluid, and phantom distant from the fluid. Visibility was determined from the contrast-to-noise ratio between the fluid and phantom, whereas streaking artifact was quantified by relative noise in the phantom. Measures were individually fit using multiple linear regression to determine an optimal contrast-to-fluid ratio for increased visualization without streaking. Contrast media- and blood-doped saline and D5W were also tested to determine whether such doping altered their electrical conductivity. RESULTS. Iohexol concentration most influenced CT number; volumetric ratios of 1:1000-1:1 produced 20 HU to over 3000 HU. CT numbers were weakly dependent on x-ray tube voltage, whereas contrast-to-noise ratio and streaking artifacts were somewhat dependent on tube output. An optimal ratio of iohexol in fluid was determined to be 1:50. There was no significant difference between the electrical impedances of doped and pure saline or D5W (p > 0.5, all cases). CONCLUSION. A 1:50 ratio of iohexol in saline or D5W provides an optimal combination of increased visibility on CT without streaking artifacts.
机译:目的。这项研究的目的是确定用于器官水切术的生理盐水中碘化造影剂的浓度和水(D5W)中5%葡萄糖的浓度,该技术用于物理分离和保护邻近热消融的组织。材料和方法。用生理盐水或D5W中的1:1000-1:1碘海醇或碘乙酸酯葡甲胺造影剂制备了总共28个样品。使用80-120 kVp和10-300 mAs通过CT对单独的样品或与均质的模仿肝脏的幻影并置的样品进行成像。从流体,背景空气,靠近流体的体模和远离流体的体模测量平均CT数和噪声。可见性是由流体与体模之间的对比噪声比确定的,而条纹伪影则通过体模中的相对噪声进行量化。使用多重线性回归对测量进行个体拟合,以确定最佳的对比度/流体比,以增加可视化效果而不会出现条纹。还测试了造影剂和血液掺杂的盐水和D5W,以确定这种掺杂是否改变了它们的电导率。结果。碘海醇浓度对CT值影响最大; 1:1000-1:1的体积比产生20 HU到3000 HU以上。 CT数量几乎不依赖于X射线管电压,而对比噪声比和条纹伪影则稍微依赖于显像管输出。确定碘海醇在液体中的最佳比例为1:50。掺杂和纯盐水或D5W的电阻抗之间没有显着差异(p> 0.5,所有情况)。结论。盐水或D5W中碘海醇的比例为1:50,可提供在CT上不可见条纹的情况下增加可见度的最佳组合。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号