首页> 外文期刊>The International journal of oral & maxillofacial implants >Is There a Need for Standardization of Tightening Force Used to Connect the Transducer for Resonance Frequency Analysis in Determining Implant Stability?
【24h】

Is There a Need for Standardization of Tightening Force Used to Connect the Transducer for Resonance Frequency Analysis in Determining Implant Stability?

机译:是否需要用于连接换能器的紧固力标准化,以便在确定植入物稳定性方面进行共振频率分析?

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: The purpose of this in vitro study was to determine the minimum placement torque required to attach the transducer (measuring peg) to the implant to provide an accurate assessment of implant stability using resonance frequency analysis. Materials and Methods: One hundred 4 x 11-mm screw-shaped titanium implants were inserted into a uniform polyurethane block with similar density to bone in a standardized surgical protocol. The implants were distributed into 10 groups, with 10 implants each (G1 to G10). In G1, the transducer was manually attached by a female operator and in G2 by a male operator using the manual connector provided by the manufacturer. For the remaining groups (G3 to G10), the transducers were placed using a connector adapted to a digital torque wrench with different torque settings: 3 Ncm (G3), 4 Ncm (G4), 5 Ncm (G5), 6 Ncm (G6), 10 Ncm (G7), 13 Ncm (G8), 17 Ncm (G9), and 20 Ncm (G10). Stability was measured for all groups using the Osstell equipment (Diagnosis of integration) and the implant stability quotient (ISQ) annotated for statistical comparison between the groups. Results: The mean +/- standard deviation ISQ values for groups G1 to G10 were 9.50 +/- 5.54, 19.05 +/- 2.67, 29.25 +/- 4.22, 26.55 +/- 5.37, 40.90 +/- 0.99, 69.60 +/- 2.41, 71.30 +/- 0.82, 71.20 +/- 1.32, 72.40 +/- 0.97, and 70.90 +/- 0.88, respectively. Statistical comparisons determined that the amplitudes of the confidence intervals, relative to the standard deviations, were lowest for groups G5, G7, G8, G9, and G10. For the means, the lowest amplitudes of the confidence intervals were observed in G6, G7, G8, G9, and G10. When checking the conjugated confidence intervals (mean and standard deviation), the results were homogenous for G7, G8, G9, and G10. When the torque of 20 Ncm was reached, the connection between the transducer and the implant failed. Conclusion: In this in vitro model experiment, transducer torques between 10 and 17 Ncm appear to be adequate for accurate measurement of implant stability, allowing more precise comparisons without damaging the prosthetic connection in the implant.
机译:目的:该体外研究的目的是确定将换能器(测量钉)连接到植入物所需的最小放置扭矩,以便使用共振频率分析提供准确评估植入物稳定性的评估。材料和方法:将一百4×11mm的螺旋形钛植入物插入均匀的聚氨酯嵌段中,具有与标准手术方案中的骨骼相似的密度。将植入物分为10组,每个植入物(G1至G10)。在G1中,使用由制造商提供的手动连接器,通过母操作器和雄性操作员手动地连接换能器。对于剩余的群体(G3至G10),使用适用于具有不同扭矩设置的数字扭矩扳手的连接器放置换能器:3ncm(g3),4ncm(g4),5 ncm(g5),6 ncm(g6 ),10nCM(G7),13 NCM(G8),17 nCM(G9)和20ncm(G10)。使用osstell设备(集成诊断)和植入稳定性(ISQ)的所有基团测量稳定性,以用于组之间的统计比较。结果:G1至G10组的平均+/-标准偏差ISQ值为9.50 +/- 5.54,19.05 +/- 2.67,29.25 +/- 4.22,26.55 +/- 5.37,40.90 +/- 0.99,69.60 + / - 2.41,71.30 +/- 0.82,71.20 +/- 1.32,72.40 +/- 0.97和70.90 +/- 0.88分别。统计比较确定相对于标准偏差的置信区间的幅度最低,对于组G5,G7,G8,G9和G10是最低的。对于该装置,在G6,G7,G8,G9和G10中观察到置信区间的最低幅度。当检查共轭置信区间(平均值和标准偏差)时,结果对于G7,G8,G9和G10均匀。达到20ncm的扭矩时,换能器与植入物之间的连接失败。结论:在这种体外模型实验中,10到17个NCM之间的换能器扭矩似乎足以准确测量植入物稳定性,从而使得更精确的比较而不会损坏植入物中的假肢连接。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号