A multimodal spectrometer for Raman scattering and near-infrared absorption measurement

机译:一种用于拉曼散射和近红外吸收测量的多峰光谱仪

获取原文
获取原文并翻译 | 示例
           

摘要

Raman and infrared (IR) spectroscopy are complementary spectroscopic techniques. However, measurement of Raman and IRspectra are commonly carried out on separate instruments. A dispersive system that enables both Raman spectroscopy and NIR spectroscopy was designed, built, and tested. The prototype system measures spectral ranges of 2600-300 cm~(-1) and 752-987 nm for Raman and NIR channels, respectively. A wavelength accuracy better than 0.6 nm and spectral resolution better than 1 nm (14.4 cm~(-1) for Raman channel) could be achieved with our configuration. The linearity of spectral response was better than 99.8%. The intensity stability of the instrument was found to be 0.7% and 0.4% for Raman and NIR channels, respectively. The performance of the instrument was evaluated using binary aqueous solutions of ethanol and ovalbumin. It was found that ethanol concentrations (2-10%) could be predicted with a root mean squared error of prediction (RMSEP) of 0.45% using Raman peak height at 882.2 cm~(-1). Quantification of ovalbumin concentration (8-16 g/L) in aqueous solutions and in denatured states yielded RMSEP values of 1.05 g/L and 0.74 g/L, respectively. Using concentration as external perturbation in two-dimensional correlation spectroscopy (2DCOS), heterospectral correlation analysis revealed the relationship between NIR and Raman spectra.
机译:拉曼和红外线(IR)光谱是互补的光谱技术。然而,RAMAN和IRSPectRA的测量通常在单独的仪器上进行。设计,构建和测试了一种可实现拉曼光谱和NIR光谱的分散系统。原型系统分别测量拉曼和NIR通道的2600-300cm〜(-1)和752-987 nm的光谱范围。可以通过我们的配置实现优于0.6nm和优于1nm(14.4cm〜(-1)的光谱分辨率的波长精度)。光谱响应的线性优于99.8%。仪器的强度稳定性分别为拉曼和NIR通道的0.7%和0.4%。使用二元水溶液的乙醇和卵烧蛋评估仪器的性能。发现乙醇浓度(2-10%)可以预测,使用拉曼峰高在882.2cm〜(-1)下的拉曼峰高的预测(Rmsep)的根平均平方误差。水溶液中卵霉浓度(8-16g / L)和变性状态的定量分别产生1.05g / L和0.74g / L的RMSEP值。在二维相关光谱(2DCOS)中使用浓度作为外部扰动,异晶谱相关分析显示了NIR和拉曼光谱之间的关系。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号