首页> 外文期刊>Tribology International >Friction force evaluation for grasping in minimally invasive surgery
【24h】

Friction force evaluation for grasping in minimally invasive surgery

机译:在微创手术中抓住摩擦力评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This study investigates direct and indirect measurement techniques for a typical laparoscopic operation in an ex vivo experiment. The direct measurement technique involves obtaining contact forces at the tissue-tool level. The indirect measurement technique involves acquiring contact forces without compromising the contact surface. The two techniques are compared in terms of the friction force. This estimation is important to ensure that grasping is stable so that tissue sliding is avoided. The results demonstrate that direct measurement is better at measuring the pinch force and estimating the friction force because of the alteration of the force center. However, integration of the tactile sensor into the tool tip increases the pull-off force by approximately 44%.
机译:本研究研究了在离体实验中典型的腹腔镜操作的直接和间接测量技术。 直接测量技术涉及在组织刀具水平处获得接触力。 间接测量技术涉及在不影响接触表面的情况下获取接触力。 在摩擦力方面比较这两种技术。 这种估计是重要的,以确保抓握是稳定的,从而避免了组织滑动。 结果表明,由于力中心的改变,直接测量更好地测量夹紧力并估计摩擦力。 然而,将触觉传感器的整合到工具尖端将拉出力增加约44%。