Evaluation of a portable ion-selective electrode meter for measuring potassium concentrations in whole blood and plasma of calves

机译:用于测量全血和犊牛血浆中钾浓度的便携式离子选择电极计的评价

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

An ion-selective electrode (ISE) handheld meter (LAQUAtwin B-731; Horiba) has recently become available for the measurement of potassium concentrations [K+] in biological fluids. The ISE meter has the potential to facilitate the diagnosis and treatment of potassium balance disorders of critically ill cattle. The objective of this study was to characterise the analytical performance of the ISE meter in a study sample of hospitalised calves with a broad range of plasma [K+]. For the purpose of the study, whole blood and plasma samples from 125 calves (age≤3months) were used for analysis. The accuracy of the meter against the reference method (indirect ISE, Cobas c 311, Roche) was assessed using Passing–Bablok regression and Bland–Altman plots.The [K+] in whole blood as measured by the ISE meter in direct mode ranged from 2.4 to 9.9mmol/L. The meter measured whole blood [K+] as 3.8% higher than plasma [K+]. Passing–Bablok regression for whole blood [K+] measured by the meter against plasma [K+] determined by indirect potentiometry revealed a linear relationship that was almost identical to the line of identity. However, the Bland–Altman plot indicated that the meter measured plasma [K+] 5.1% lower than the reference method. This result was consistent with analytical differences of direct and indirect ISE methods in respect to variation in the plasma protein concentration. In conclusion, the LAQUAtwin B-731 meter provides an accurate, rapid and low-cost tool for the diagnosis of potassium derangements in critically ill calves, particularly when whole blood samples are analysed.
机译:最近可用于测量生物流体中的钾浓度[k +]的离子选择电极(ISE)手持式电磁表(Saquatwin B-731; Horiba)。 ISE仪表有可能促进危重牛的钾平衡障碍的诊断和治疗。本研究的目的是表征ISE仪表的分析性能,在具有宽范围的等离子体[k +]的住院犊牛的研究样本中。出于该研究的目的,使用125只小牛(≤3个月)的全血和血浆样品用于分析。使用通过Bablok回归和Bland-Altman Plots评估仪表对参考方法的准确性(间接ISE,COBAS C 311,Roche)。通过ISE计以直接模式测量的全血中的[k +]范围2.4至9.9mmol / l。该仪表将全血[k +]高于比血浆[k +]高出3.8%。通过间接电位测量的仪表测量的全血[k +]的通过 - Bablok回归互相确定的血浆[k +]显示与身份线几乎相同的线性关系。然而,Bland-Altman图表明仪表测量的等离子体[k +]比参考方法低5.1%。该结果与关于血浆蛋白浓度变化的直接和间接ISE方法的分析差异一致。总之,萨基特武文B-731仪表提供了一种准确,快速,低成本的工具,用于诊断危重病的牛犊中的钾紊乱,特别是当分析全血样品时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号