Recently, surgical site infections due to non-tuberculous mycobacteria (NTM) have been linked to heater-cooler unit contamination. The European Centre for Disease Prevention and Control and manufacturers now recommend the use of hydrogen peroxide in filtered water to fill heater-cooler unit tanks. After implementation of these measures in our hospital, heater-cooler units became heavily contaminated by opportunistic waterborne pathogens such as Pseudomonas aeruginosa and Stenotrophomonas maltophilia. No NTM were detected but fast-growing resistant bacteria could impair their detection. The efficiency of hydrogen peroxide and chlorhexidine-alcohol was compared in situ. Chlorhexidine-alcohol treatment stopped waterborne pathogen contamination and NTM were not cultured whereas their detection efficiency was probably improved. (C) 2018 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.
机译:最近,由于非结核性分枝杆菌(NTM)引起的手术部位感染与加热器冷却器单元污染有关。 欧洲疾病预防和控制中心现在建议使用过氧化氢在过滤水中填充加热器冷却器单元罐。 在我们医院实施了这些措施后,加热器单元由机会水上的水性病原体污染,例如假单胞菌铜绿假单胞菌和牙科科菌疗法麦芽菌菌。 未检测到NTM,但耐快的抗性细菌可能会损害其检测。 比较原位的过氧化氢和氯己定醇的效率。 氯己定醇处理停止了水性病原体污染,NTM没有培养,而它们的检测效率可能得到改善。 (c)2018年医疗保健感染协会。 elsevier有限公司出版。保留所有权利。