Comparison of positive pressure ventilation devices in a newborn manikin

机译:新生儿型人体模型正压通风装置的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Objective: To compare tidal volume (V-T) delivery and ventilation rate between devices for positive pressure ventilation (PPV) during newborn resuscitation.Methods: Neonatal resuscitation program providers (n=25) delivered PPV to a newborn manikin in a randomized order with: a self-inflating bag (SIB), a disposable T-piece, a non-disposable T-piece, a stand-alone infant resuscitation system T-piece and the volume-controlled prototype Next Step(TM) device (KM Medical). All T-pieces used a peak inflation pressure of 20cmH(2)O and a 5cmH(2)O positive end-expiratory pressure (PEEP). The SIB neither had a PEEP valve nor manometer. The Next Step(TM) had a 5cmH(2)O PEEP valve. The participants aimed to deliver a 5mL/kg V-T (rate 40-60min(-1)) for 1min with each device and each of three compliances (0.5, 1.0 and 2.0mL/cmH(2)O). V-T and ventilation rate were compared between devices and compliance levels (ANOVA)Results: All devices, except the Next Step(TM) delivered a 4-5mL/kg V-T at the low compliance, but three- to four-fold that of the target at the higher compliance levels. The Next Step(TM) delivered a V-T close to target at all compliance levels. The ventilation rate was within 40-60min(-1) with all devices and compliance levels.Conclusions: Routinely used ventilation devices for newborn resuscitation can triple intended V-T and requires further investigation.
机译:目的:在新生儿复苏期间比较潮气量(VT)输送和通风率进行正压通风(PPV)。方法:新生儿复苏计划提供者(n = 25)以随机顺序向新生儿人体潜水到新生儿:a自充气袋(SIB),一次性T形件,一种非一次性T型T型,一个独立的婴儿复苏系统T型件和音量控制的原型下一步(TM)装置(KM医学)。所有T型件使用20cmH(2)o和5cmH(2)℃的峰值膨胀压力(PEEP)。 SIB既不有窥视阀也没有压力计。下一步(TM)具有5CMH(2)O PEEP阀。参与者旨在提供5ml / kg V​​-T(速率40-60min(-1)),其中每个设备,三个顺应性中的每一个(0.5,1.0和2.0ml / cmH(2)O)。在设备和符合性水平(ANOVA)结果(ANOVA)结果之间进行了VT和通风率:所有设备,除了下一步(TM)以低遵从性输送4-5ml / kg V​​T,但三到四倍的目标在较高的合规级别。下一步(TM)在所有合规级别都传递到目标的V-T。通风率在40-60min(-1)范围内,具有所有设备和合规性水平。结论:常规使用用于新生儿复苏的通风装置可以三重预期V-T,需要进一步调查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号