Purpose: To compare the foveal avascular zone (FAZ) area measurements produced by different optical coherence tomography angiography (OCTA). Methods: Healthy enrolled volunteers underwent OCTA using 2 different devices: Spectralis HRA+OCTA (Heidelberg Engineering, Heidelberg, Germany) and RS-3000 Advance (Nidek, Gamagori, Japan). Two graders measured FAZ in both superficial (SCP) and deep (DCP) retinal capillary plexuses. The SCP and DCP en face images were visualized automatically segmenting 2 separate slabs defined by the arbitrary segmentation lines created by the software of each OCT device. One grader repeated each measure twice. Results: Fifty-nine eyes were included. The mean FAZ was 0.33 +/- 0.09 mm(2) at the SCP and 0.57 +/- 0.17 mm(2) at the DCP measured with RS-3000 versus 0.30 +/- 0.08 and 0.35 +/- 0.08 mm(2), respectively, measured with Spectralis. The measurements of the 2 devices were significantly different (p 0.0001). The intraoperator agreement was excellent at the SCP (intraclass correlation coefficient, ICC: 0.97 with Spectralis and 0.96 with RS-3000). At the DCP, it was good with Spectralis and fair with RS-3000 (ICC: 0.85 and 0.64, respectively). The interoperator agreement was excellent for Spectralis and good for RS-3000 at the SCP (ICC: 0.97 and 0.93, respectively). It was good at the DCP with both devices (ICC: 0.74 with RS-3000 and 0.81 with Spectralis). Conclusions: FAZ measurements obtained with different OCTA devices differ. These findings should be considered in followup studies of patients with retinal vascular diseases. (C) 2018 S. Karger AG, Basel
机译:目的:比较不同光学相干断层造影血管造影(OctA)产生的软体扶手区(FAZ)区域测量。方法:健康注册志愿者使用2种不同的设备进行Octa:Spectralis HRA + Octa(海德堡工程,海德堡,德国)和RS-3000进步(Nidek,Gamagori,日本)。两位成绩在表面(SCP)和深(DCP)视网膜毛细经膜中测量FAZ。 SCP和DCP en面部图像被自动分割由由每个OCT设备的软件创建的任意分段线定义的单独的平板。一分级器两次重复两次。结果:包括五十九次眼睛。 SCP的平均Faz为0.33 +/- 0.09 mm(2),并在DCP下测量为0.57 +/- 0.17 mm(2),测量为RS-3000与0.30 +/- 0.08和0.35 +/- 0.08 mm(2)分别用光谱测量。 2器件的测量值显着不同(P <0.0001)。 SCP的史内术协议非常出色(腹部相关系数,ICC:0.97,光谱和0.96,RS-3000)。在DCP,它良好的光谱和公平,RS-3000(ICC:0.85和0.64分别)。互操作器协议非常适合Spectralis,并且在SCP(ICC:0.97和0.93分别)对RS-3000良好。它是擅长DCP的两个设备(ICC:0.74,RS-3000和Spectralis 0.81)。结论:用不同Octa器件获得的FAZ测量值不同。应考虑这些发现在视网膜血管疾病的患者的后续研究中。 (c)2018年S. Karger AG,巴塞尔

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号