Image quality and dose analysis for a PA chest X-ray: Comparison between AEC mode acquisition and manual mode using the 10 kVp 'rule'

机译:PA胸X射线的图像质量和剂量分析:使用10 kVP'规则'的AEC模式采集与手动模式的比较

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To compare the image quality and effective dose applying the 10 kVp rule with manual mode acquisition and AEC mode in PA chest X-ray. Method: 68 images (with and without lesions) were acquired using an anthropomorphic chest phantom using a Wolverson Arcoma X-ray unit. These images were compared against a reference image using the 2 alternative forced choice (2AFC) method. The effective dose (E) was calculated using PCXMC software using the exposure parameters and the DAP. The exposure index (IgM provided by Agfa systems) was recorded. Results: Exposure time decreases more when applying the 10 kVp rule with manual mode (50%-28%) when compared with automatic mode (36%-23%). Statistical differences for E between several ionization chambers' combinations for AEC mode were found (p = 0.002). £ is lower when using only the right AEC ionization chamber. Considering the image quality there are no statistical differences (p = 0.348) between the different ionization chambers' combinations for AEC mode for images with no lesions. Considering IgM values, it was demonstrated that they were higher when the AEC mode was used compared to the manual mode. It was also observed that IgM values obtained with AEC mode increased as kVp value went up. The image quality scores did not demonstrate statistical significant differences (p = 0.343) for the images with lesions comparing manual with AEC mode. Conclusion: In general the £ is lower when manual mode is used. By using the right AEC ionising chamber under the lung the £ will be the lowest in comparison to other ionising chambers. The use of the 10 kVp rule did not affect the visibility of the lesions or image quality.
机译:目的:将图像质量和有效剂量进行比较,在PA胸X射线中使用手动模式采集和AEC模式应用10 kVP规则。方法:使用Wolverson Arcoma X射线单元使用拟人胸部幻影获得68图像(具有和没有病变)。使用2替代强制选择(2AFC)方法将这些图像与参考图像进行比较。使用曝光参数和DAP使用PCXMC软件计算有效剂量(E)。记录曝光指数(AGFA系统提供的IGM)。结果:当与自动模式相比,在使用手动模式(50%-28%)时,曝光时间会降低更多(50%-28%)(36%-23%)。发现了几种电离室之间的e的统计差异(p = 0.002)。仅使用右侧AEC电离室时较低。考虑到图像质量在没有病变的图像的AEC模式的不同电离室的组合之间没有统计差(P = 0.348)。考虑到IGM值,证明它们与手动模式相比使用AEC模式时更高。还观察到,随着KVP值上升时,使用AEC模式获得的IGM值增加。图像质量分数未对具有AEC模式进行比较手册的图像的图像没有演示统计显着差异(P = 0.343)。结论:通常,当使用手动模式时,通常£较低。通过在肺下的右侧AEC电离室,与其他电离室相比,£将是最低的。 10 KVP规则的使用不影响病变或图像质量的可见性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号