Development of a computer simulation technique for low-dose chest radiographs: a phantom study

机译:低剂量胸部射线照相计算机仿真技术的开发:幻影研究

获取原文
获取原文并翻译 | 示例
           

摘要

The present study aimed to develop a simple computer simulation method of low-dose radiographs based on a radiograph acquired at a clinical-dose level. A chest phantom was used for the development of this method. In this method, a simulated low-dose image was obtained from a clinical-dose image using an input-output characteristic curve of a flat panel detector and noise metrics of the standard deviation (SD) and noise power spectrum. We applied this method for low-dose images of a chest phantom to evaluate the simulation accuracy. The noise SDs were compared between the simulated and real images corresponding to 1/2, 1/4, and 1/8 of clinical doses. The relative error of noise SDs in the chest phantom images was less than 3%. Therefore, we believe that the proposed simulation method has the potential to be useful for determination of the optimal exposure condition in chest radiography to reduce patients' exposure dose.
机译:本研究旨在基于在临床剂量级别获得的射线照片的射线照片开发一种简单的低剂量射线照片的简单计算机模拟方法。 胸部幻影用于开发这种方法。 在该方法中,使用标准偏差(SD)和噪声功率谱的平板检测器和噪声度量的输入 - 输出特性曲线从临床剂量图像获得模拟的低剂量图像。 我们将这种方法应用于胸部幻影的低剂量图像,以评估模拟精度。 在对应于1/2,1 / 4和1/8的临床剂量的模拟和真实图像之间比较噪声SDS。 胸部幻像图像中的噪声SDS的相对误差小于3%。 因此,我们认为,所提出的仿真方法具有可用于测定胸部射线照相中最佳暴露条件以减少患者的曝光剂量。