Evaluation of quantitative I-123 and I-131 SPECT with Monte Carlo-based down-scatter compensation

机译:定量I-123和I-131 SPECT与基于蒙特卡罗的下散射补偿的评价

获取原文
获取原文并翻译 | 示例
           

摘要

Objective Quantitative I-123 and I-131 single-photon emission computed tomography (SPECT) is hampered by down-scatter from the high-energy peaks. This paper presents a down-scatter compensation method, where down-scatter generated in the patient and gamma camera collimator and detector is modelled using Monte Carlo simulation in the ordered subsets expectation maximization SPECT reconstruction algorithm. Materials and methods The new down-scatter compensation method was compared with conventional triple energy window (TEW) scatter compensation and Gaussian convolution-based forced detection Monte Carlo methods. The comparison was made with the NEMA-IEC phantom using six spherical inserts (diameters from 10 to 37 mm) and a lung compartment. The phantom was filled with I-123 and I-131 solutions to known sphere-to-background concentration ratios. Spherical volumes of interest with the same diameter as the inserts were drawn on the images, and recovery coefficients for the spheres were calculated in addition to lung-to-background ratio. Results The new down-scatter compensation method provided higher recovery coefficients than the TEW scatter compensation or Gaussian convolution-based forced detection Monte Carlo algorithm for both isotopes. Background activity concentration could be accurately estimated with the new down-scatter compensation method and with the TEW scatter compensation, whereas activity concentration of the spheres was severely underestimated even with the new method. Conclusion Down-scatter compensation with Monte Carlo-simulation effectively reduces down-scatter effects in I-123 and I-131 SPECT imaging.
机译:目的定量I-123和I-131单光子发射电流扫描(SPECT)被从高能峰的落后阻碍。本文介绍了一个下散射补偿方法,其中在患者和伽马相机准直器和检测器中产生的下散射使用蒙特卡罗模拟在有序的子集期望最大化SPECT重建算法中建模。材料和方法与传统三维能量窗口(TEW)散射补偿和基于高斯卷积的强制检测蒙特卡罗方法进行比较新的下散射补偿方法。使用六个球形插入物(直径为10至37mm)和肺隔室,用NEMA-IEC Phantom进行比较。幻影被I-123和I-131解填充到已知的球形到背景浓度比。在图像上绘制与插入件相同的直径的球形体积,除了肺到背景比外,还计算用于球体的回收系数。结果新的脱散散射补偿方法提供了比TEW散射补偿或基于高斯卷积的强制检测蒙特卡罗算法的恢复系数更高。背景技术即使用新方法也可以准确地估计新的下散射补偿方法和仪表散射补偿。结论利用蒙特卡罗模拟的下散射补偿有效地降低了I-123和I-131 SPECT成像中的下散射效果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号