An ex vivo comparison of cuffed endotracheal tubes and self-sealing baffled endotracheal tubes

机译:袖口气管管和自密封挡板气管内管的exvivo比较

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

AIMSTo compare cuffed silicone endotracheal tubes with self-sealing baffled silicone endotracheal tubes in an ex vivo canine tracheal model, to determine whether the tubes differed in their maintenance of a seal and their effectiveness in removing fluid from the tracheal lumen, and whether the self-sealing endotracheal tubes would release pressure when a closed anaesthetic circuit reached 30cm H2O.METHODSTwelve cadaver tracheae were randomly selected to be intubated with either cuffed or self-sealing baffled endotracheal tubes. To test tracheal seal efficacy, the tracheae were positioned vertically, 5mL of water was instilled proximal to the tube cuff or baffles and they were monitored for leakage at 0, 15, 30, 45, and 60 minutes, recording the total volume leaked. To test fluid removal at extubation, the tracheae were intubated, then 5.6g of 60% barium sulphate suspension was instilled proximal to the seal and left undisturbed for 5 minutes. The tubes were then extubated, with the cuffed endotracheal tubes being partially deflated before extubation, and the amount of barium recovered was weighed. These procedures were repeated on the same tracheae using the other endotracheal tubes. To test whether self-sealing baffled endotracheal tubes would release pressure at 30cm H2O, the tracheae were intubated, connected to an anaesthetic machine and pressurised to 30cm H2O for 5 minutes and then 50cm H2O for 5 minutes. Release of pressure was defined as a decrease in pressure within the closed anaesthetic circuit.RESULTSMore cuffed (7/12) than baffled (0/12) endotracheal tubes leaked water after 60 minutes (p=0.016). The mean amount of barium removed by the self-sealing baffled endotracheal tubes (4.9 (95% CI=3.8-4.4) g) was greater than that removed by the partially deflated cuffed endotracheal tubes (0.4 (95% CI=0.14-0.66) g) (p0.001). None of the self-sealing baffled endotracheal tubes released pressure at 30cm H2O in a closed anaesthetic circuit.CONCLUSIONS AND CLINICAL RELEVANCESelf-sealing baffled endotracheal tubes were more effective than cuffed endotracheal tubes at both preventing fluid leakage at the tracheal seal and removing fluid from the lumen of the trachea in cadaver tracheae. However they did not release pressure when the closed-circuit system was at 30cm H2O in a canine cadaver model. The self-sealing baffled endotracheal tubes may be a suitable substitute for cuffed endotracheal tubes.
机译:Aimsto比较带有自密封堵塞硅氧烷气管管的袖扣硅胶气管管,以确定管道是否在其维持密封和从气管流体中除去流体的效果,以及自我 - 当闭合麻醉回路达到30cm H2O时,密封气管管将释放压力。随机选择用袖口或自密封挡板的气管内部外伤管随机选择加管尸体尸体气管。为了测试气管密封功效,将气管垂直定位,将5mL水浸泡在管箍或挡板上,在0,15,30,45和60分钟中监测它们,记录泄漏的总体积。为了在拔管处进行流体去除,将气管预热,然后将5.6g 60%的硫酸钡悬浮液浸泡在密封件近侧并保持不受干扰5分钟。然后拔管管,用拔管前部分放气的袖带气管膜,称重钡的量。使用其他气管膜管在同一轮气管上重复这些程序。为了测试自密封的挡板气管膜是否会在30cm H 2 O下释放压力,加管装载气管,连接到麻醉剂,并将其加压至30cm H 2 O 5分钟,然后50cm H 2 O。压力释放被定义为闭合麻醉电路内的压力降低。卷曲(7/12)比挡板(0/12)气管插管在60分钟后泄漏水(P = 0.016)。通过自密封挡板气管膜(4.9(95%Ci = 3.8-4.4)克上除去的平均钡量大于通过部分放气的柔软气管膜管除去(0.4(95%Ci = 0.14-0.66) g)(p <0.001)。没有一个自密封的挡板气管管在闭合麻醉电路中释放了30cm H 2 O的压力。结论和临床相关性密封的困扰的气管内膜管在防气管密封液中的液体泄漏并从中移除流体气管的内腔在尸体装饰尸体。然而,当闭路系统在犬尸体模型中闭合电路系统在30厘米H2O时,它们没有释放压力。自密封挡板气管膜管可以是适当的袖干气管管替代品。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号