A modification to the standard Winkler's method (WM) for measurement of dissolved oxygen (DO) concentration in aqueous samples enabled DO measurements in 1 mL sample volume. However, advanced statistical analysis indicated that the proposed method (PM) showed a small positive bias, giving results that were on average about 6% higher than corresponding values measured using the WM. In this study, DO measurements were carried out in sample volumes ranging from 2 mL to 10 mL using the PM. Statistical analysis of these results indicate that the bias reported in the PM could be eliminated (i.e., limited to a +/- 2% range) using sample volumes of 7 mL or higher. Use of a larger sample volume also enables lowering the least count of the PM to 0.1 mg/L O-2 as compared to 0.25 mg/L O-2 reported previously. (C) 2017 Elsevier Ltd. All rights reserved.
机译:对水性样品中的溶解氧(DO)浓度测量的标准Winkler的方法(WM)的修改使得在1mL样品体积中进行测量。 然而,高级统计分析表明,所提出的方法(PM)显示出小的阳性偏压,给出平均比使用WM测量的相应值高约6%的结果。 在该研究中,使用PM的样品体积中进行测量,使用PM测量从2mL至10ml。 这些结果的统计分析表明PM中报道的偏差可以使用7ml或更高的样品体积来消除(即,限制为+/- 2%的范围)。 与先前报道的0.25mg / L O-2相比,使用较大的样品体积也能够降低PM至0.1mg / L O-2的最小计数。 (c)2017 Elsevier Ltd.保留所有权利。