Introduction: Merging robotics with laser eye surgery could enhance precision, repeatability and automation. During some eye laser procedures the patient is awake, thus eye stabilization is desired to avoid movements that could affect the treatment. Material and methods: The ESPRESSO platform has a two-stage actuation system to position a stabilization tool on the eye, a proximity sensing unit to monitor the stabilization tool position, and a sensing unit to monitor the pressure exerted on the eye. The platform is tested in-vitro and ex-vivo with clinicians. A maximum pressure to be exerted on the eye is defined with expert ophthalmic surgeons to be 22 mmHg: physiological intraocular pressure (IOP) range is 10-21 mmHg. This pressure corresponds to a force of 0.3 N. Results: The necessary contact force to have eye fixation (according to the clinicians' feedback) is evaluated: maximum values resulted always below 0.3 N. A maximum IOP increase of 4.67 mmHg is observed, that is a slight variation with respect to the performance of other platforms (IOP elevations up to 328 mmHg). Conclusion: Design and initial assessment of the platform is presented. Eye stabilization is performed without exceeding the critical contact force value and causing large/sudden IOP increases.
机译:简介:用激光眼科手术合并机器人可以提高精度,可重复性和自动化。在一些眼睛激光程序期间,患者唤醒,因此需要眼睛稳定来避免可能影响治疗的运动。材料和方法:浓缩咖啡平台具有两级致动系统,可在眼睛上定位稳定工具,一个接近感测单元来监测稳定工具位置,以及监测眼睛上施加的压力的传感单元。该平台与临床医生在体外和前体内测试。在眼睛上施加的最大压力是用专家眼科外科医生定义为22 mmHg:生理眼内压(IOP)范围为10-21mmHg。该压力对应于0.3 n的力。结果:评价必要的接触力(根据临床医生的反馈):最大值导致始终低于0.3 n。观察到最大IOP增加4.67mmHg,即对于其他平台的性能(IOP高度高达328mmHg)是略有变化的。结论:提出了对平台的设计和初始评估。在不超过临界接触力值的情况下进行眼睛稳定,导致大/突然的IOP增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号