【24h】

Simultaneous determination of Co, Fe, Ni and K with HR CS GFAAS

机译:同时测定CO,Fe,Ni和K与HR CS GFAAs

获取原文
获取原文并翻译 | 示例
           

摘要

In this work, a method for simultaneous determination of four different elements, Co, Fe, Ni and K with high resolution continuum source graphite furnace atomic absorption spectrometry was developed. For this purpose less sensitive lines of four elements were selected (344.364 nm for Co, 344.388 nm for Fe, 344.626 nm for Ni and 344.641 nm for K). The pyrolysis and atomization temperatures were of 1000 and 2000 degrees C, respectively, and no chemical modifier was necessary. Calibration against aqueous standards lead to good agreement between certified or informed values and the determined values, at a statistical confidence level of 95%; recovery tests were performed for real samples, resulting in recoveries ranging from 90 to 110%. Limit of quantification for 0.050 mg L-1 Co, 0.173 mg L-1 Fe, 0.267 mg L-1 Ni and 0.629 mg L-1 K have been obtained, which are adequate for the purpose.
机译:在这项工作中,开发了一种用于同时测定具有高分辨率连续源石墨炉原子吸收光谱法的四种不同元素,CO,Fe,Ni和k的方法。 为此目的,选择了四种元素的敏感线(对于Co,Fe,344.388nm,对于k的344.626nm,344.626nm,k)。 热解和雾化温度分别为1000和2000℃,并且不需要化学改性剂。 抵御水平标准的校准导致认证或知情人员之间的良好一致性和确定的价值观,统计置信水平为95%; 对真实样品进行恢复试验,导致回收率范围为90至110%。 为0.050mg L-1的定量限制,得到0.173mg L-1 Fe,0.267mg L-1 Ni和0.629mg L-1 k,这是为了目的是足够的。

著录项