【24h】

Determination of the thermodynamic temperature between 236 K and 430 K from speed of sound measurements in helium

机译:从氦气中的声音测量速度测定236 k和430 k之间的热力学温度

获取原文
获取原文并翻译 | 示例
           

摘要

We report speed of sound measurements in helium at 273.16 K and at eight temperatures in the range between 236K and 430 K. These results determine the difference (T - T-90) between the thermodynamic temperature T and its approximation T-90 by the International Temperature Scale of 1990 (ITS-90). The uncertainty of our measurements of (T - T-90) spans between a minimum of 0.25 mK near 247 K and a maximum of 0.89 mK at the freezing point of indium (429.75 K) with comparable contributions from the uncertainty of our acoustic determination of T and from the uncertainty of our laboratory realization of ITS-90. On the overlapping temperature ranges these results are consistent with other recent acoustic determinations of (T - T-90). We also present evidence that (T - T-90) can be determined with comparably small uncertainties by the alternative, time-saving procedure of measuring the speed-of-sound in helium using only a single, judiciously-chosen, pressure on each isotherm.
机译:我们在273.16 k氦气中报告氦气中的声音测量速度,并且在236k和430k的范围内。这些结果决定了国际热力学温度T及其近似T-90之间的差异(T-T-90) 1990年的温度尺度(ITS-90)。 我们测量的(T-T-90)测量的不确定性在最低0.25 mk接近247 k附近,最大为0.89 mk,在铟的冷冻点(429.75 k),具有来自我们声学测定的不确定性的可比贡献 T来自我们-90的实验室实现的不确定性。 在重叠温度范围内,这些结果与(T-T-90)的其他最近的声学测定一致。 我们还提出了(T-T-90)可以通过替代,节省氦气在每个等温线上测量氦气速度的替代,节省时间的替代,节省时间的步骤来确定(T-T-90)的证据。 。