Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation

机译:流行病学调查期间采用振动血压测量的质量控制与验证

获取原文
获取原文并翻译 | 示例
           

摘要

This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist.Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (standard device). Two percent of subjects enrolled in an epidemiological investigation were systematically sampled in this study. Intraclass correlation coefficient (ICC) was used to evaluate measurement reliability, paired t-test was used to evaluate trueness, and linear regression was used for calibration. The Association for the Advancement of Medical Instrumentation (AAMI) standards and British Hypertension Society (BHS) protocols were used for validation quality assessment.Both mercury sphygmomanometer (standard device) and oscillometric BP monitor (test device) displayed high reliability. A significant difference in systolic blood pressure (SBP) was observed between devices. SBP calibration was achieved by using an effective linear regression model (B=0.803 and constant=19.592, P<0.001). The calibrated model was corroborated by verification samples (P=0.120) and was found to pass AAMI standards and BHS protocol requirements.Calibrated SBP measurements from the Omron HBP-1300 device were valid. Use of a combination of statistical methods, such as ICC for reliability assessment as well as paired t-test for trueness evaluation can be used to validate data from the oscillometric BP monitors.
机译:本研究旨在验证由示波器BP监视器测量的血压(BP)值,并且如果存在差异,则寻求可能的校准方法。使用示波器BP监视器(OMRON HBP-1300)与汞血压计(标准装置)确定了未进行BP测量结果。 )。在本研究中系统地抽样了两种参加流行病学调查的受试者的百分之一。脑内相关系数(ICC)用于评估测量可靠性,使用配对的T检验来评估真实性,并且使用线性回归用于校准。医疗仪表(AAMI)标准和英国高血压协会(BHS)协议的协会用于验证质量评估。汞血压计(标准装置)和示波器BP监视器(测试装置)显示出高可靠性。在器件之间观察到收缩压(SBP)的显着差异。通过使用有效的线性回归模型实现了SBP校准(B = 0.803和常数= 19.592,P <0.001)。通过验证样本(P = 0.120)证实了校准模型,并发现通过AAMI标准和BHS协议要求。来自欧姆龙HBP-1300设备的alibrated SBP测量有效。使用统计方法的组合,例如ICC进行可靠性评估以及用于改革评估的配对T检验,可用于验证来自示波器测量BP监视器的数据。

著录项

  • 来源
    《Medicine.》 |2015年第37期|共7页
  • 作者单位

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Key Lab Tumor Epidemiol Henan Prov Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

    Zhengzhou Univ Coll Publ Hlth Dept Epidemiol &

    Hlth Stat 100 Kexue Ave Zhengzhou 450001 Henan;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 医药、卫生;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号