AIM OF THE STUDY: The goal of this study was to compare the attachment of stapes prostheses with differently shaped loops to the long process of the incus. METHOD: Gold, steel/Teflon, platinum/Teflon, and two different titanium stapes prostheses were inserted in 30 specially prepared temporal bones by three experienced surgeons using the Fisch technique with the McGee crimper and straight alligator forceps for the crimping of the loops. RESULTS: In all prostheses, a sufficiently firm attachment of the long process of the incus was achieved. The band-shaped loops showed a better contact with the incus than did the wire loops. However, the broad spiral-shaped loops led to a loss of the perpendicular axis of the piston to the long incus process. CONCLUSION: The geometry of the loop affects the final length of the piston in the vestibule and its angle to the long process of the incus.
机译:研究的目的:这项研究的目的是比较形状不同的带的pro骨假体与cus骨的长过程的连接。方法:由三位经验丰富的外科医生使用Fisch技术,将McGee压接钳和直形鳄鱼钳用于环的压接,将金,钢/铁氟龙,铂/铁氟龙和两种不同的钛骨假体插入30个特殊准备的颞骨中。结果:在所有假体中,都可以充分牢固地固定the骨的长进程。带状环显示出比金属环更好的与the骨的接触。但是,宽的螺旋形环导致活塞的垂直轴失去了较长的针距。结论:套环的几何形状会影响前庭中活塞的最终长度以及其与砧骨长过程的角度。