首页> 外文期刊>Medical dosimetry: official journal of the American Association of Medical Dosimetrists >Use of electronic portal images to evaluate setup error and intra-fraction motion during free-breathing breast IMRT treatment
【24h】

Use of electronic portal images to evaluate setup error and intra-fraction motion during free-breathing breast IMRT treatment

机译:在自由呼吸乳房IMRT治疗期间使用电子门户图像评估设置误差和分数内运动

获取原文
获取原文并翻译 | 示例
           

摘要

Before delivering of intensity-modulated radiotherapy, kilo-voltage image-guidance radiotherapy is widely used in setup error correction and monitoring intra-fraction motion effectively. Accordingly, this study proposes and tests an image integration technique for observing intra-fraction motion during beam delivery, with the wider objective of reducing both image-guidance time and the dose delivered to normal breast tissue. The study sample comprised 33 female patients with breast cancer, and 241 sets of portal images acquired using a VARIAN aSi-1000 electronic portal imaging device. Motion amplitudes and vectors were collected and calculated separately by two senior therapists. The setup error in 3 axes was computed for every fraction, with average shifting for lateral, longitudinal and vertical direction was -0.3-mm +/- 0.5, -0.1-mm +/- 0.5 and -0.6-mm +/- 1.6, with the average vector of setup error being 2.9-mm +/- 1.4. The average intra-fraction motion for vertical direction was (A: -0.1-mm +/- 1.0; B: -0.0 +/- 1.1), for longitudinal was (A: -0.4-mm +/- 1.7; B: 2.0 +/- 1.1), and for lateral direction was (A: 0.3-mm +/- 1.3; B: 0.2 +/- 1.8). The average intra-fraction vector was 2.9-mm +/- 1.3 for therapist A, and 3.4-mm +/- 1.8 for therapist B. Offline Review commercial software was utilized for setup error and motion analysis, and data analysis and reliability testing were conducted with statistical package of the social sciences. Pearson correlations between the two therapists was moderate (0.59, p 0.01), and the Cohen's kappa value for inter rater agreement between different evaluators was fair in the anterior-posterior direction (0.25, p 0.01), with slight agreement in other two directions and vectors. The study presented efficient and dose reduction method to evaluate setup error and intra-fraction motion during breast intensity-modulated radiotherapy treatment. (C) 2018 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
机译:在递送强度调制的放射疗法之前,千电压图像引导放射疗法广泛用于设置纠错和有效监测分数内运动。因此,本研究提出并测试了用于在光束输送期间观察分数内运动的图像集成技术,具有更广泛的目的,即减少图像引导时间和输送到正常乳房组织的剂量。该研究样本包含33例女性乳腺癌患者,并使用Varian ASI-1000电子门户成像装置获得的241套门户图像。由两位高级治疗师分开收集和计算运动幅度和载体。为每个分数计算3轴的设置误差,横向,纵向和垂直方向的平均换档为-0.3mm +/- 0.5,-0.1-mm +/- 0.5和-0.6-mm +/- 1.6,使用2.9 mm +/- 1.4的设置错误的平均矢量。垂直方向的平均级分钟运动是(A:0.1-mm +/- 1.0; b:-0.0 +/- 1.1),用于纵向(A:0.4-mm +/- 1.7; B:2.0 +/- 1.1)和横向(A:0.3-mm +/- 1.3; B:0.2 +/- 1.8)。对于治疗师A.治疗师B的平均分数载体为2.9mm +/- 1.3,用于治疗师B.离线审查商业软件用于设置误差和运动分析,以及数据分析和可靠性测试与社会科学的统计包进行。两位治疗师之间的Pearson相关性适中(0.59,P 0.01),并且在不同评估员之间的帧间协议中的COHEN的Kappa值在前后方向上公平(0.25,P 0.01),略有协议其他两个方向和向量。该研究提出了高效和剂量的减少方法,以评估乳房强度调制的放射治疗过程中的设置误差和分数内运动。 (c)2018年美国医疗剂量分子协会。由elsevier Inc.保留所有权利发布。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号