Multi‐gradient‐echo myelin water fraction imaging: Comparison to the multi‐echo‐spin‐echo technique

机译:多梯度回声髓鞘水分成像:与多回波 - 旋转回波技术的比较

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Purpose Myelin water fraction (MWF) mapping based on multi‐gradient recalled‐echo (MGRE) imaging has been proposed as an alternative to the conventional multi‐echo‐spin‐echo (MESE) approach. In this work, we performed a comparative study of MESE and MGRE‐derived MWFs in the same subject group. Methods MESE and MGRE data were acquired in 12 healthy volunteers at 3T. T 2 * decay curves were corrected for the effects of field inhomogeneities and multicomponent analysis of T 2 ? and T 2 * signals was performed using non‐negative least‐squares fitting. Results When comparing MGRE and MESE‐MWFs across volunteers, no significant differences were observed between average values in WM, deep GM (dGM), and cortical GM (cGM) that were (14?±?3%), (6?±?2%), and (8?±?2%) for MGRE, and (13?±?2%), (6?±?1%), and (7?±?1%), respectively, for MESE. The MGRE and MESE‐MWFs showed a strong correlation (r 2 ?=?0.84) and Bland‐Altman analysis revealed a small positive bias of (0.8?±?1.6%) (absolute difference) for the MGRE‐MWF. Conclusion Overall, we observed excellent agreement between the two techniques. The small positive bias of the MGRE‐MWF is thought to be a consequence of its potentially reduced sensitivity to water exchange effects, compared to the MESE‐MWF. This work suggests that with careful correction for the effects of field inhomogeneities, MGRE‐MWF imaging is a promising alternative to the MESE approach. Magn Reson Med 79:1439–1446, 2018. ? 2017 International Society for Magnetic Resonance in Medicine.
机译:目的,基于多梯度回忆回波(MGRE)成像的目的髓鞘水分(MWF)映射被提出为传统的多回波 - 旋转回波(MESE)方法的替代方案。在这项工作中,我们对同一主题群体进行了对Mese和MGRE衍生的MWF的比较研究。方法在3T的12个健康志愿者中获得MESE和MGRE数据。 T 2 *衰减曲线被纠正了现场不均匀性的影响和T 2的多组分分析?使用非负至少方块配件进行T 2 *信号。结果在志愿者跨行员比较MGRE和MESE-MWF时,在WM,深丙克(DGM)和皮质GM(CGM)的平均值之间没有显着差异(14?±3%),(6?±?用于MGRE的2%)和(8?±2%),(13?±2%),(6?±1%),和(7?±1%),用于MESE。 MGRE和MESE-MWF显示出强烈的相关性(R 2?= 0.84),并且Bland-Altman分析显示了MGRE-MWF的(0.8?±1.6%)的小正偏差(0.8?±1.6%)。结论总体而言,我们观察了两种技术之间的良好一致。与MESE-MWF相比,MGRE-MWF的小阳性偏差被认为是其对水交换效应的敏感性降低的结果。这项工作表明,通过对现场不均匀性的影响进行仔细校正,MGRE-MWF成像是对MESE方法的有希望的替代品。 Magn Reson Med 79:1439-1446,2018 2017年医学磁共振的国际社会。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号